Title: Evaluation of a Variable-Aperture Full-Ring SPECT System using Large-Area Pixelated CZT Modules: A Simulation Study for Brain SPECT Applications

Abstract

Single photon emission computed tomography (SPECT) scanner using cadmium zinc telluride (CZT) offers improved imaging capability over conventional NaI (Tl)-based SPECT scanner. We aim to demonstrate a full-ring SPECT system design with eight large-area CZT detectors that can be used for a broad spectrum of SPECT radiopharmaceuticals and provides higher sensitivity and better spatial resolution than those of conventional NaI (Tl)-based gamma cameras. A newly-designed full-ring SPECT system is composed of 8 large-area CZT cameras (128 mm × 179.2 mm effective area) that can be independently swiveled around their own axes of rotation independently and can have radial motion for varying aperture sizes that can be adapted to different sizes of imaging volume. Extended projection data were generated by conjoining projections of two adjacent detectors to overcome the limited field-of-view (FOV) by each CZT camera. Using Monte Carlo simulations, we evaluated this new system design with digital phantoms including a Derenzo hot-rod phantom and a Zubal brain phantom. Comparison of performance metrics such as spatial resolution, sensitivity, contrast-to-noise ratio (CNR), and contrast-recovery ratio were made between our design and a conventional SPECT scanner. The proposed scanner could result in up to about 3 times faster in acquisition time over conventionalmore » scan time at same acquisition time per step. The spatial resolution of our proposed scanner was similar or better to that of the conventional scanner, and there was significant performance improvement over the conventional scanner particularly in sensitivity (approximately 4 times). Overall, we successfully reconstructed the phantom image for both 99mTc-based perfusion and 123I-based dopamine transporter (DaT) brain studies simulated for our new design. In particular, the striatal/background contrast-recovery ratio in 3-to-1 reference ratio was over the 0.8 for the 123I-based DaT study. In conclusion, we demonstrated the potential of our new full-ring CZT SPECT design, showing improved performance metrics such as improved system sensitivity and CNR while maintaining other important imaging parameters such as spatial resolution.« less

Authors:
 [1];  [1];  [2];  [2];  [3];  [3];  [4];  [4]
  1. Univ. of California, San Francisco, CA (United States)
  2. Brookhaven National Lab. (BNL), Upton, NY (United States)
  3. Shanghai Jiao Tong Univ. (China)
  4. Univ. of California, San Francisco, CA (United States); Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
Publication Date:
Research Org.:
Brookhaven National Lab. (BNL), Upton, NY (United States)
Sponsoring Org.:
USDOE National Nuclear Security Administration (NNSA), Office of Nonproliferation and Verification Research and Development; National Inst. of Biomedical Imaging and Bioengineering; National Heart, Lung, and Blood Inst.
OSTI Identifier:
1770726
Report Number(s):
BNL-213700-2021-JAAM
Journal ID: ISSN 0094-2405
Grant/Contract Number:  
SC0012704; R01EB026331; R01EB012965; R01HL135490
Resource Type:
Accepted Manuscript
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 48; Journal Issue: 5; Journal ID: ISSN 0094-2405
Publisher:
American Association of Physicists in Medicine
Country of Publication:
United States
Language:
English
Subject:
36 MATERIALS SCIENCE; SPECT; CZT detector; Full ring; Brain SPECT; Monte Carlo simulation

Citation Formats

Huh, Yoonsuk, Yang, Jaewon, Dim, Odera U., Cui, Yonggang, Tao, Weijie, Huang, Qiu, Gullberg, Grant T., and Seo, Youngho. Evaluation of a Variable-Aperture Full-Ring SPECT System using Large-Area Pixelated CZT Modules: A Simulation Study for Brain SPECT Applications. United States: N. p., 2021. Web. doi:10.1002/mp.14836.
Huh, Yoonsuk, Yang, Jaewon, Dim, Odera U., Cui, Yonggang, Tao, Weijie, Huang, Qiu, Gullberg, Grant T., & Seo, Youngho. Evaluation of a Variable-Aperture Full-Ring SPECT System using Large-Area Pixelated CZT Modules: A Simulation Study for Brain SPECT Applications. United States. https://doi.org/10.1002/mp.14836
Huh, Yoonsuk, Yang, Jaewon, Dim, Odera U., Cui, Yonggang, Tao, Weijie, Huang, Qiu, Gullberg, Grant T., and Seo, Youngho. Thu . "Evaluation of a Variable-Aperture Full-Ring SPECT System using Large-Area Pixelated CZT Modules: A Simulation Study for Brain SPECT Applications". United States. https://doi.org/10.1002/mp.14836. https://www.osti.gov/servlets/purl/1770726.
@article{osti_1770726,
title = {Evaluation of a Variable-Aperture Full-Ring SPECT System using Large-Area Pixelated CZT Modules: A Simulation Study for Brain SPECT Applications},
author = {Huh, Yoonsuk and Yang, Jaewon and Dim, Odera U. and Cui, Yonggang and Tao, Weijie and Huang, Qiu and Gullberg, Grant T. and Seo, Youngho},
abstractNote = {Single photon emission computed tomography (SPECT) scanner using cadmium zinc telluride (CZT) offers improved imaging capability over conventional NaI (Tl)-based SPECT scanner. We aim to demonstrate a full-ring SPECT system design with eight large-area CZT detectors that can be used for a broad spectrum of SPECT radiopharmaceuticals and provides higher sensitivity and better spatial resolution than those of conventional NaI (Tl)-based gamma cameras. A newly-designed full-ring SPECT system is composed of 8 large-area CZT cameras (128 mm × 179.2 mm effective area) that can be independently swiveled around their own axes of rotation independently and can have radial motion for varying aperture sizes that can be adapted to different sizes of imaging volume. Extended projection data were generated by conjoining projections of two adjacent detectors to overcome the limited field-of-view (FOV) by each CZT camera. Using Monte Carlo simulations, we evaluated this new system design with digital phantoms including a Derenzo hot-rod phantom and a Zubal brain phantom. Comparison of performance metrics such as spatial resolution, sensitivity, contrast-to-noise ratio (CNR), and contrast-recovery ratio were made between our design and a conventional SPECT scanner. The proposed scanner could result in up to about 3 times faster in acquisition time over conventional scan time at same acquisition time per step. The spatial resolution of our proposed scanner was similar or better to that of the conventional scanner, and there was significant performance improvement over the conventional scanner particularly in sensitivity (approximately 4 times). Overall, we successfully reconstructed the phantom image for both 99mTc-based perfusion and 123I-based dopamine transporter (DaT) brain studies simulated for our new design. In particular, the striatal/background contrast-recovery ratio in 3-to-1 reference ratio was over the 0.8 for the 123I-based DaT study. In conclusion, we demonstrated the potential of our new full-ring CZT SPECT design, showing improved performance metrics such as improved system sensitivity and CNR while maintaining other important imaging parameters such as spatial resolution.},
doi = {10.1002/mp.14836},
journal = {Medical Physics},
number = 5,
volume = 48,
place = {United States},
year = {Thu Mar 11 00:00:00 EST 2021},
month = {Thu Mar 11 00:00:00 EST 2021}
}

Works referenced in this record:

An energy-optimized collimator design for a CZT-based SPECT camera
journal, January 2016

  • Weng, Fenghua; Bagchi, Srijeeta; Zan, Yunlong
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 806
  • DOI: 10.1016/j.nima.2015.09.115

Simulations of a Multipinhole SPECT Collimator for Clinical Dopamine Transporter (DAT) Imaging
journal, September 2018

  • Konik, Arda; De Beenhouwer, Jan; Mukherjee, Joyeeta M.
  • IEEE Transactions on Radiation and Plasma Medical Sciences, Vol. 2, Issue 5
  • DOI: 10.1109/TRPMS.2018.2831208

GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy
journal, January 2011


Monte Carlo-based scatter correction considering the tailing effect of a CdTe detector for dual-isotope brain SPECT imaging
journal, July 2016

  • Suzuki, Atsuro; Takeuchi, Wataru; Ueno, Yuichiro
  • Biomedical Physics & Engineering Express, Vol. 2, Issue 4
  • DOI: 10.1088/2057-1976/2/4/045010

SNM Practice Guideline for Dopamine Transporter Imaging with 123I-Ioflupane SPECT 1.0
journal, December 2011

  • Djang, D. S. W.; Janssen, M. J. R.; Bohnen, N.
  • Journal of Nuclear Medicine, Vol. 53, Issue 1
  • DOI: 10.2967/jnumed.111.100784

ACR–ACNM Practice Parameter for the Performance of Dopamine Transporter (DaT) Single Photon Emission Computed Tomography (SPECT) Imaging for Movement Disorders
journal, January 2017

  • Subramaniam, Rathan M.; Frey, Kirk A.; Hunt, Christopher H.
  • Clinical Nuclear Medicine, Vol. 42, Issue 11
  • DOI: 10.1097/RLU.0000000000001815

Simulation-based evaluation and optimization of a new CdZnTe gamma-camera architecture (HiSens)
journal, April 2010

  • Robert, Charlotte; Montémont, Guillaume; Rebuffel, Véronique
  • Physics in Medicine and Biology, Vol. 55, Issue 9
  • DOI: 10.1088/0031-9155/55/9/019

Advances in clinical molecular imaging instrumentation
journal, January 2018

  • Hutton, Brian F.; Erlandsson, Kjell; Thielemans, Kris
  • Clinical and Translational Imaging, Vol. 6, Issue 1
  • DOI: 10.1007/s40336-018-0264-0

Accurate Monte Carlo modelling of the back compartments of SPECT cameras
journal, November 2010


Image Reconstruction is a New Frontier of Machine Learning
journal, June 2018

  • Wang, Ge; Ye, Jong Chu; Mueller, Klaus
  • IEEE Transactions on Medical Imaging, Vol. 37, Issue 6
  • DOI: 10.1109/TMI.2018.2833635

First assessment of simultaneous dual isotope (123I/99mTc) cardiac SPECT on two different CZT cameras: A phantom study
journal, March 2017

  • Blaire, Tanguy; Bailliez, Alban; Ben Bouallegue, Fayçal
  • Journal of Nuclear Cardiology, Vol. 25, Issue 5
  • DOI: 10.1007/s12350-017-0841-z

Validation of the Monte Carlo simulator GATE for indium-111 imaging
journal, June 2005


Recent advances in cardiac SPECT instrumentation and imaging methods
journal, March 2019


Computerized three-dimensional segmented human anatomy
journal, February 1994

  • Zubal, I. George; Harrell, Charles R.; Smith, Eileen O.
  • Medical Physics, Vol. 21, Issue 2
  • DOI: 10.1118/1.597290

Quantitative capabilities of four state-of-the-art SPECT-CT cameras
journal, January 2012


The role of hybrid SPECT-CT in oncology: current and emerging clinical applications
journal, March 2008


Monte Carlo Simulations of the GE Discovery Alcyone CZT SPECT Systems
journal, June 2015

  • Pretorius, P. Hendrik; Liu, Chi; Fan, Peng
  • IEEE Transactions on Nuclear Science, Vol. 62, Issue 3
  • DOI: 10.1109/TNS.2015.2433533

Optimized sampling for high resolution multi-pinhole brain SPECT with stationary detectors
journal, January 2020

  • Chen, Yuan; Goorden, Marlies C.; Vastenhouw, Brendan
  • Physics in Medicine & Biology, Vol. 65, Issue 1
  • DOI: 10.1088/1361-6560/ab5bc6

STIR: software for tomographic image reconstruction release 2
journal, January 2012

  • Thielemans, Kris; Tsoumpas, Charalampos; Mustafovic, Sanida
  • Physics in Medicine and Biology, Vol. 57, Issue 4
  • DOI: 10.1088/0031-9155/57/4/867

Preliminary Investigation of Axial and Angular Sampling in Multi-Pinhole AdaptiSPECT-C with XCAT Phantoms
conference, October 2017

  • Zeraatkar, Navid; Kalluri, Kesava S.; Konik, Arda
  • 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
  • DOI: 10.1109/NSSMIC.2017.8532835

Integration of advanced 3D SPECT modeling into the open-source STIR framework: SPECT reconstruction with STIR
journal, August 2013

  • Fuster, Berta Marti; Falcon, Carles; Tsoumpas, Charalampos
  • Medical Physics, Vol. 40, Issue 9
  • DOI: 10.1118/1.4816676

Japanese multicenter database of healthy controls for [123I]FP-CIT SPECT
journal, February 2018

  • Matsuda, Hiroshi; Murata, Miho; Mukai, Yohei
  • European Journal of Nuclear Medicine and Molecular Imaging, Vol. 45, Issue 8
  • DOI: 10.1007/s00259-018-3976-5

Feasibility study of a novel general purpose CZT-based digital SPECT camera: initial clinical results
journal, March 2018


CZT device with improved sensitivity for medical imaging and homeland security applications
conference, September 2009

  • Chen, H.; Awadalla, S. A.; Marthandam, P.
  • SPIE Optical Engineering + Applications, SPIE Proceedings
  • DOI: 10.1117/12.828514

Performance evaluation of a novel high performance pinhole array detector module using NEMA NU-4 image quality phantom for four head SPECT Imaging
conference, March 2015

  • Rahman, Tasneem; Tahtali, Murat; Pickering, Mark R.
  • SPIE Medical Imaging, SPIE Proceedings
  • DOI: 10.1117/12.2082365

Advances in imaging instrumentation for nuclear cardiology
journal, July 2017


A four-pixel matched collimator for high-sensitivity SPECT imaging
journal, March 2013

  • Suzuki, Atsuro; Takeuchi, Wataru; Ishitsu, Takafumi
  • Physics in Medicine and Biology, Vol. 58, Issue 7
  • DOI: 10.1088/0031-9155/58/7/2199

Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology
journal, April 2009

  • Erlandsson, Kjell; Kacperski, Krzysztof; van Gramberg, Dean
  • Physics in Medicine and Biology, Vol. 54, Issue 9
  • DOI: 10.1088/0031-9155/54/9/003

Generative Image Inpainting with Contextual Attention
conference, June 2018

  • Yu, Jiahui; Lin, Zhe; Yang, Jimei
  • 2018 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR)
  • DOI: 10.1109/CVPR.2018.00577

The subresolution DaTSCAN phantom: a cost-effective, flexible alternative to traditional phantom technology
journal, January 2018


European multicentre database of healthy controls for [123I]FP-CIT SPECT (ENC-DAT): age-related effects, gender differences and evaluation of different methods of analysis
journal, November 2012

  • Varrone, Andrea; Dickson, John C.; Tossici-Bolt, Livia
  • European Journal of Nuclear Medicine and Molecular Imaging, Vol. 40, Issue 2
  • DOI: 10.1007/s00259-012-2276-8

Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging
journal, October 2003

  • Pareto, Deborah; Cot, Albert; Pav�a, Javier
  • European Journal of Nuclear Medicine and Molecular Imaging, Vol. 30, Issue 10
  • DOI: 10.1007/s00259-003-1229-7

Including Anatomical and Functional Information in MC Simulation of PET and SPECT Brain Studies. Brain-VISET: A Voxel-Based Iterative Method
journal, October 2014

  • Marti-Fuster, Berta; Esteban, Oscar; Thielemans, Kris
  • IEEE Transactions on Medical Imaging, Vol. 33, Issue 10
  • DOI: 10.1109/TMI.2014.2326041

CNN-based PET sinogram repair to mitigate defective block detectors
journal, December 2019


Design of a CZT based breast SPECT system
journal, June 1998

  • Singh, M.; Mumcuoglu, E.
  • IEEE Transactions on Nuclear Science, Vol. 45, Issue 3
  • DOI: 10.1109/23.681995