Improving FAIMS Sensitivity using a Planar Geometry with Slit Interfaces

Journal Article · · Journal of the American Society for Mass Spectrometry, 20(9):1768-1774
Differential mobility spectrometry or field asymmetric waveform ion mobility spectrometry (FAIMS) is gaining broad acceptance for analyses of gas-phase ions, especially in conjunction with largely orthogonal separation methods such as mass spectrometry (MS) and/or conventional (drift tube) ion mobility spectrometry. In FAIMS, ions are filtered while passing through a gap between two electrodes that may have planar or curved (in particular, cylindrical) geometry. Despite substantial inherent advantages of the planar configuration and its universal acceptance in stand-alone FAIMS devices, commercial FAIMS/MS systems have employed curved FAIMS geometries that could be interfaced to MS more effectively. Here we report a new planar (p-) FAIMS design with slit-shaped entrance and exit apertures that substantially increase ion transmission in and out of the analyzer. The front slit interface effectively couples p-FAIMS to multi-emitter electrospray ionization (ESI) sources, improving greatly the ion current introduced to the device. The back slit interface increases the transmission of ribbon-shaped ion beams output by the p-FAIMS to downstream stages such as a MS. Overall, the ion signal in ESI/FAIMS/MS analyses is raised by over an order of magnitude without affecting the FAIMS resolution.
Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (US), Environmental Molecular Sciences Laboratory (EMSL)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
965561
Report Number(s):
PNNL-SA-65387; 20496; 400412000
Journal Information:
Journal of the American Society for Mass Spectrometry, 20(9):1768-1774, Journal Name: Journal of the American Society for Mass Spectrometry, 20(9):1768-1774 Journal Issue: 9 Vol. 20; ISSN JAMSEF; ISSN 1044-0305
Country of Publication:
United States
Language:
English

Similar Records

High-Resolution Field Asymmetric Waveform Ion Mobility Spectrometry Using New Planar Geometry Analyzers
Journal Article · Thu Jun 01 00:00:00 EDT 2006 · Analytical Chemistry, 78(11):3706-3714 · OSTI ID:885457

Hooked differential mobility spectrometry apparatus and method therefore
Patent · Mon Feb 16 23:00:00 EST 2009 · OSTI ID:988295

Scaling of the Resolving Power and Sensitivity for Planar FAIMS and Mobility-Based Discrimination in Flow- and Field- Driven Analyzers
Journal Article · Mon Sep 10 00:00:00 EDT 2007 · Journal of the American Society for Mass Spectrometry, 18(9):1672–1681 · OSTI ID:917958