ENCAPSULATED SURFACE-BARRIER PARTICLE DETECTORS SOME METHODS AND TECHNIQUES

Technical Report ·
DOI:https://doi.org/10.2172/4794108· OSTI ID:4794108
Techniques are described in detail that have been developed for the manufacture of encapsulated surface-barrier particle detectors. The devices manufactured to date have found use as counters and spectrometers for fission fragments, alpha particles, and low-energy electrons from radioactive nuclei. Devices, 50 mm/sup 2/ in area, have a resolution for 6.1-Mev alpha particles of 26 kev at room temperature and 18 kev at 78 deg K. Resolution for 1.1-Mev electrons has been measured as 7.5 kev at 78 deg K for devices 25 mm/sup 2/ in area. The silicon used has been n-type, grown with a 111 orientation, and purchased in the form of ingots about 2 cm in diameter; the resistivity has ranged from 100 to 3000 ohm-cm. (auth)
Research Organization:
Argonne National Lab., Ill.
DOE Contract Number:
W-31109-ENG-38
NSA Number:
NSA-16-014901
OSTI ID:
4794108
Report Number(s):
ANL-6526
Country of Publication:
United States
Language:
English

Similar Records

Silicon Surface-Barrier Nuclear Particle Spectrometer
Journal Article · Wed Jun 01 00:00:00 EDT 1960 · IRE Transactions on Nuclear Science · OSTI ID:4140455

IMPROVEMENTS IN ENCAPSULATED SILICON JUNCTION ALPHA DETECTORS
Journal Article · Sat Dec 31 23:00:00 EST 1960 · IRE Trans. Nuclear Sci. · OSTI ID:4100271

THE POSSIBILITIES OF USING SOLID-STATE SILICON DETECTORS IN NUCLEAR STUDIES
Journal Article · Fri Nov 30 23:00:00 EST 1962 · Izvestiya Akademii Nauk, Seriya Fizicheskaya (U.S.S.R.) For English translation see Bull. Acad. Sci. USSR, Phys. Ser. (Engl. Transl.) · OSTI ID:4727308