Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

SCINTILLATION SPECTROMETER WITH THE ELIMINATION OF THE COMPTON BACKGROUND

Journal Article · · Krajowe Symp. Zastosowan Izotopow Tech., Rogow, Pol., 1960
OSTI ID:4722519
A method is suggested for suppressing the Compton background in gamma spectrum measurements using a scintillation spectrometer with two NaI(Tl) crystals connected in coincidence. The block diagram of the experimental system is given, and results of measurements of the gamma spectra of Co/sup 60/, Cs/sup 13/, and (Co/sup 60/ + Cs/sup 137/) sources are report ed. The gamma spectrum curve peaks correspond to the maximum energy of recoil electrons. It is reported that the resolution of the lines proved to be 20% less accurate than that of the photolines. (Referativnyy zhurnal, Fizika, No. 10, 1962)
Research Organization:
Originating Research Org. not identified
NSA Number:
NSA-17-023760
OSTI ID:
4722519
Journal Information:
Krajowe Symp. Zastosowan Izotopow Tech., Rogow, Pol., 1960, Journal Name: Krajowe Symp. Zastosowan Izotopow Tech., Rogow, Pol., 1960 Vol. Vol: No. 4
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

SCINTILLATION BETA-SPECTROMETER
Journal Article · Sat Dec 31 23:00:00 EST 1960 · Radioktivn. Izlucheniya i Metody ikh Issled., Akad. Nauk Latv. SSR, Inst. Fiz. · OSTI ID:4065166

A SCINTILLATOR FOR THE RECORDING OF SLOW NEUTRONS
Journal Article · Sat Dec 31 23:00:00 EST 1960 · Krajowe Symp. Zastosowan Izotopow Tech., Rogow, Pol., 1960 · OSTI ID:4722518

GAMMA SPECTROMETER WITH ORGANIC SCINTILLATOR CAPABLE OF TIME RESOLVING SELECTION OF $gamma$ RADIATION
Journal Article · Tue Feb 28 23:00:00 EST 1961 · Pribory i Tekh. Ekspt. · OSTI ID:4840309