THE PROPERTIES OF CaSO$sub 4$: Mn THERMOLUMINESCENCE DOSIMETERS

Technical Report ·
OSTI ID:4694746
The properties of the radiothermoluminescence of CaSO/sub 4/: Mn, used for dosimetry of gamma and roentgen radiation, were investigated. The light yield of the luminophor was determined to 1.5% for 1 Mev gamma radiation. The dependence of the thermoluminescence light sum on the exposure, the exposure rate, and the exposure time can qualitatively be described by the first order process model, modified by a broad energy distribution of the electron trap depths. Some applications are discussed. The cheapness of the dosimeters, the convenient read-out, and the broad range of measurable exposures suggest that CaSO/sub 4/: Mn can be a valuable complement to other dosimetry systems. (auth)
Research Organization:
Aktiebolaget Atomenergi, Stockholm
NSA Number:
NSA-17-028820
OSTI ID:
4694746
Report Number(s):
AE-109
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

CaSO$sub 4$(Mn) THERMOLUMINESCENCE DOSIMETERS FOR SMALL DOSES OF GAMMA RADIATION
Journal Article · Mon Oct 01 00:00:00 EDT 1962 · Rev. Sci. Instr. · OSTI ID:4764551

MEASUREMENTS OF SMALL EXPOSURES OF GAMMA RADIATION WITH CaSO$sub 4$:Mn RADIOTHERMOLUMINESCENCE
Technical Report · Thu Aug 01 00:00:00 EDT 1963 · OSTI ID:4652749

A THERMOLUMINESCENT DOSIMETER
Journal Article · Wed Aug 01 00:00:00 EDT 1962 · Kernenergie (East Germany) · OSTI ID:4746123