MEASUREMENTS OF SMALL EXPOSURES OF GAMMA RADIATION WITH CaSO$sub 4$:Mn RADIOTHERMOLUMINESCENCE

Technical Report ·
OSTI ID:4652749
A system for measurements of small exposures of gamma radiation using CaSO/sub 4/: Mn thermoluminescence was developed. The construction and performance of a read-out apparatus is described as well as the construction and characteristics of a simple dosimeter. The reproducibility of the method at various exposures is estimated. 20 mu r of 1 Mev gamma radiation can be measured with a reproducibility within plus or minus 50% (standard deviation). (auth)
Research Organization:
Aktiebolaget Atomenergi, Stockholm
NSA Number:
NSA-17-036141
OSTI ID:
4652749
Report Number(s):
AE-118
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

CaSO$sub 4$(Mn) THERMOLUMINESCENCE DOSIMETERS FOR SMALL DOSES OF GAMMA RADIATION
Journal Article · Mon Oct 01 00:00:00 EDT 1962 · Rev. Sci. Instr. · OSTI ID:4764551

THE PROPERTIES OF CaSO$sub 4$: Mn THERMOLUMINESCENCE DOSIMETERS
Technical Report · Sat Jun 01 00:00:00 EDT 1963 · OSTI ID:4694746

A THERMOLUMINESCENT DOSIMETER
Journal Article · Wed Aug 01 00:00:00 EDT 1962 · Kernenergie (East Germany) · OSTI ID:4746123