Measurement of the spectra of fission fragments against the background of large alpha particle fluxes

Journal Article · · Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1063- 1065
OSTI ID:4343203
Translated from Prib. Tekn. Eksp.; 16: No. 4, 69-71(1973). A simple electronic circuit is described that is designed for the spectrometry of fission fragments by means of a semiconductor detector. The circuit performs amplification, selection, and stretching of pulses from fragments and allows their amplitude to be measured for an alpha-particle flux intensity to the detector equal to 10/sup 6/ sec/sup -1/ and a fragment intensity that is 10/sup 8/ to 10/ sup 9/ times smaller. Data are presented on the radiation immunity of p-- n detectors made of low-ohm Si in fluxes of alpha -particles and neutrons. (auth)
Research Organization:
Originating Research Org. not identified
NSA Number:
NSA-29-018480
OSTI ID:
4343203
Journal Information:
Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1063- 1065, Journal Name: Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1063- 1065; ISSN INETA
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

ZnS + B scintillation counter: a threshold detector for a pulsed flux of fast neutrons
Journal Article · Sun Dec 31 23:00:00 EST 1972 · Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1080- 1082 · OSTI ID:4369737

Analysis of methods for measuring neutron flux in the presence of gamma radiation
Journal Article · Sun Dec 31 23:00:00 EST 1972 · Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1032- 1034 · OSTI ID:4319399

Gi(Li) detector with separate sensing regions and certain of its applications in neutron activation analysis
Journal Article · Sun Dec 31 23:00:00 EST 1972 · Instrum. Exp. Tech. (USSR) (Engl. Transl.), v. 16, no. 4, pp. 1066- 1070 · OSTI ID:4343249