Sensitive Single Channel Pulse-Height Analyzer
Journal Article
·
· Review of Scientific Instruments
A stable sensitive single channel pulse-height analyzer is described, in which 6BE6 pentagrid tubes paralleled by vacuum diodes are used as linear dis4- to 40-mv amplitude, and it can be operated directly by a scintillation detector. The dead time is approximately 6 mu sec, and the stability of the instrument over a period of several days is better than 1%.
- Research Organization:
- Weizmann Inst. of Science, Rehovoth, Israel
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-12-017566
- OSTI ID:
- 4300778
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 9 Vol. 29; ISSN 0034-6748; ISSN RSINAK
- Publisher:
- American Institute of Physics (AIP)
- Country of Publication:
- Country unknown/Code not available
- Language:
- English