Title: Advances in x-ray computed microtomography at the NSLS

Conference ·
OSTI ID:674877

The X-Ray Computed Microtomography workstation at beamline X27A at the NSLS has been utilized by scientists from a broad range of disciplines from industrial materials processing to environmental science. The most recent applications are presented here as well as a description of the facility that has evolved to accommodate a wide variety of materials and sample sizes. One of the most exciting new developments reported here resulted from a pursuit of faster reconstruction techniques. A Fast Filtered Back Transform (FFBT) reconstruction program has been developed and implemented, that is based on a refinement of the gridding algorithm first developed for use with radio astronomical data. This program has reduced the reconstruction time to 8.5 sec for a 929 x 929 pixel{sup 2} slice on an R10,000 CPU, more than 8x reduction compared with the Filtered Back-Projection method.

Research Organization:
Brookhaven National Lab. (BNL), Upton, NY (United States)
Sponsoring Organization:
USDOE Office of Energy Research, Washington, DC (United States)
DOE Contract Number:
AC02-98CH10886
OSTI ID:
674877
Report Number(s):
BNL-65832; CONF-980851-; ON: DE99000942; TRN: 99:001161
Resource Relation:
Conference: 47. Denver X-ray conference on advances in X-ray analysis, Colorado Springs, CO (United States), 3-7 Aug 1998; Other Information: PBD: Aug 1998