Title: ADVANCES IN X-RAY COMPUTED MICROTOMOGRAPHY AT THE NSLS.

Conference ·
OSTI ID:10389

The X-Ray Computed Microtomography workstation at beamline X27A at the NSLS has been utilized by scientists from a broad range of disciplines from industrial materials processing to environmental science. The most recent applications are presented here as well as a description of the facility that has evolved to accommodate a wide variety of materials and sample sizes. One of the most exciting new developments reported here resulted from a pursuit of faster reconstruction techniques. A Fast Filtered Back Transform (FFBT) reconstruction program has been developed and implemented, that is based on a refinement of the ''gridding'' algorithm first developed for use with radio astronomical data. This program has reduced the reconstruction time to 8.5 sec for a 929 x 929 pixel{sup 2} slice on an R10,000 CPU, more than 8x reduction compared with the Filtered Back-Projection method.

Research Organization:
Brookhaven National Lab. (BNL), Upton, NY (United States)
Sponsoring Organization:
USDOE Office of Energy Research (ER) (US)
DOE Contract Number:
AC02-98CH10886
OSTI ID:
10389
Report Number(s):
BNL-65832; KC0204011; R&D Project: LS1; KC0204011; TRN: US0103537
Resource Relation:
Conference: ADVANCES IN X-RAY ANALYSIS, FORTY SEVENTH DENVER X-RAY CONFERENCE, VOLUME 42, COLORADO SPRINGS, CO (US), 08/03/1998--08/07/1998; Other Information: PBD: 7 Aug 1998
Country of Publication:
United States
Language:
English