Validation and Verification of the Correct Operation of the Thermo Fisher Scientific’s Genesys 10S spectrometer used in ANC122 and NF-ANC122
- Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
In December of 2015, the old qualified Prim UV-Vis instrument used in the ANC122 and NFANC122 procedures failed its monthly calibration check. Attempts to get the system to pass by cleaning flow cells, replacing the light bulb and realigning the light path failed to get sustained reproducible results from the system. As it could no longer pass QA/QC requirements the decision to take it out of service was made. To replace the system, a previously purchased Thermo Fisher Scientific UV-VIS spectrometer (model Genesys 10S; Serial #: 2L5R059137) was retrieved from storage. This report shows that the system is performing as required and meeting all QC requirements listed in ANC122 and NF-ANC122
- Research Organization:
- Los Alamos National Laboratory (LANL), Los Alamos, NM (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC52-06NA25396
- OSTI ID:
- 1239065
- Report Number(s):
- LA-UR-16-21031
- Country of Publication:
- United States
- Language:
- English
Similar Records
Long-Term Carbon Injection Field Test for 90% Mercury Removal for a PRB Unit a Spray Dryer and Fabric Filter
Photo–thermo Catalytic Oxidation over a TiO2-WO3-Supported Platinum Catalyst
Photo–thermo Catalytic Oxidation over a TiO 2 ‐WO 3 ‐Supported Platinum Catalyst
Technical Report
·
Thu Apr 30 00:00:00 EDT 2009
·
OSTI ID:1239065
Photo–thermo Catalytic Oxidation over a TiO2-WO3-Supported Platinum Catalyst
Journal Article
·
Sun Apr 05 00:00:00 EDT 2020
· Angewandte Chemie
·
OSTI ID:1239065
+9 more
Photo–thermo Catalytic Oxidation over a TiO 2 ‐WO 3 ‐Supported Platinum Catalyst
Journal Article
·
Fri May 29 00:00:00 EDT 2020
· Angewandte Chemie (International Edition)
·
OSTI ID:1239065
+9 more