The first synchrotron infrared beamlines at the Advanced Light Source: Spectromicroscopy and fast timing
Two recently commissioned infrared beamlines on the 1.4 bending magnet port at the Advanced Light Source, LBNL, are described. Using a synchrotron as an IR source provides three primary advantages: increased brightness, very fast light pulses, and enhanced far-IR flux. The considerable brightness advantage manifests itself most beneficially when performing spectroscopy on a microscopic length scale. Beamline (BL) 1.4.3 is a dedicated FTIR spectromicroscopy beamline, where a diffraction-limited spot size using the synchrotron source is utilized. BL 1.4.2 consists of a vacuum FTIR bench with a wide spectral range and step-scan capability. This BL makes use of the pulsed nature of the synchrotron light as well as the far-IR flux. Fast timing is demonstrated by observing the pulses from the electron bunch storage pattern at the ALS. Results from several experiments from both IR beamlines will be presented as an overview of the IR research currently being done at the ALS.
- Research Organization:
- Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
- Sponsoring Organization:
- USDOE Director, Office of Science. Office of Basic Energy Studies. Division of Materials Sciences (US)
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 764347
- Report Number(s):
- LBNL-44208; R&D Project: 458121; TRN: US0005197
- Resource Relation:
- Conference: 4th International Conference on Low Energy Electrodynamics in Solids, LEES '99, Pecs (HU), 06/21/1999--06/25/1999; Other Information: PBD: 1 Sep 1999
- Country of Publication:
- United States
- Language:
- English