Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A Josephson array voltage standard at 10 V

Journal Article · · IEEE Electron Device Lett.; (United States)

The technology of Josephson voltage standards has been extended to an array of 14 184 junctions which is capable of generating over 150 000 quantized voltage levels spanning the range from -12 to +12V. This makes possible the direct calibration of 10-V Zener reference standards without the use of a voltage divider.

Research Organization:
Electromagnetic Technology Div., National Bureau of Standards, Boulder, CO (US)
OSTI ID:
7246342
Journal Information:
IEEE Electron Device Lett.; (United States), Journal Name: IEEE Electron Device Lett.; (United States) Vol. EDL-8:10; ISSN EDLED
Country of Publication:
United States
Language:
English

Similar Records

A 10-V Josephson voltage standard
Journal Article · Fri Mar 31 23:00:00 EST 1989 · IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA) · OSTI ID:5169684

Comparisons of 10 V Josephson array voltage standards between the BNM/LCIE and the BIPM
Conference · Fri Dec 30 23:00:00 EST 1994 · OSTI ID:125686

A transfer device for 10 V Josephson array measurements
Conference · Fri Dec 30 23:00:00 EST 1994 · OSTI ID:125665