Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A transfer device for 10 V Josephson array measurements

Conference ·
OSTI ID:125665
;  [1]
  1. Bureau International des Poids et Mesures, Sevres (France)

To improve measurements of 10 V Zener based standards, we have designed a new transfer voltage source for use with 10 V Josephson arrays. The transfer voltage is compared, in turn, with that of the Zener standard and of the array. The use of two separate circuits should help to solve the problem of common ground connections and prevent noise from the Zener standard entering the array circuit. This transfer device can also be used for 10 V Josephson array comparisons.