Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Gain variation with gold coating thickness in pulsed microchannel plate striplines

Technical Report ·
OSTI ID:6958967
We have built and fielded gated stripline microchannel plate (MCP) detectors since 1984. We have always coated the input (stripline) side with 1 ..mu..m of gold, and the output or ground plane side with 1.5 ..mu..m of gold. It now appears that this input-side coating thickness is excessive, significantly reducing the gain obtainable from the detector. The strong gain reduction is correlated with slight closing up of the 12-..mu..m openings of the microchannels in the input side of the MCP. 4 refs., 7 figs.
Research Organization:
Lawrence Livermore National Lab., CA (USA)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6958967
Report Number(s):
UCID-21321; ON: DE88012447
Country of Publication:
United States
Language:
English