A neutron dose detector with rem response to 1 GeV
Conference
·
· Transactions of the American Nuclear Society; (United States)
OSTI ID:6947703
- Lawrence Berkeley Lab., CA (United States)
- Los Alamos National Lab., NM (United States)
The limitation of current rem meters, which do not measure neutron dose equivalents above [approx] 15 MeV, is a serious problem at high-energy accelerator facilities, where a much wider range of neutron energies exist. The purpose of this work was to measure the response of a modified Andersson-Braun (A-B) rem meter to neutron energies up to 1 GeV. The modifications to the standard A-B rem meter were based on the experimental results of Pb(n,xn) reactions.
- OSTI ID:
- 6947703
- Report Number(s):
- CONF-921102--
- Conference Information:
- Journal Name: Transactions of the American Nuclear Society; (United States) Journal Volume: 66
- Country of Publication:
- United States
- Language:
- English