A neutron dose detector with REM response to 1 GeV
Conference
·
OSTI ID:10184265
- Lawrence Berkeley Lab., CA (United States)
- Los Alamos National Lab., NM (United States)
The limitation of current remmeters, which do not measure neutron dose equivalents above about 15 MeV, is a serious problem at high-energy accelerator facilities, where a much wider range of neutron energies exist. The purpose of this work was to measure the response of a modified Anderson-Braun (A-B) remmeter to neutron energies up to 1 GeV. The modifications to the standard A-B remmeter were based on the experimental results of Pb(n,xn) reactions.
- Research Organization:
- Lawrence Berkeley Lab., CA (United States)
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 10184265
- Report Number(s):
- LBL--32347; CONF-921102--20; ON: DE92041203