Detector, shielding and geometric design factors for a high-resolution PET system
- California Univ., Los Angeles, CA (USA). Div. of Nuclear Medicine
The authors have evaluated the resolution, efficiency and scatter rejection on a new high resolution PET system designed for animal studies which is based on a 2-D modular detector system. A digital positioning system was evaluated by testing different encoding methods. Tungsten inter-plane septa of different thicknesses and geometries were evaluated by Monte Carlo simulations and experiments. The detector system consists of a 6 {times} 8 array of BGO crystals coupled to 2 dual photomultiplier tubes (PMTs). The crystals are 3.5 mm wide with 4 mm spacing transaxially and are 6.25 mm long with 6.75 mm spacing axially. PMT outputs are digitized and Anger camera type logic is used to determine the X and Y location of the scintillation event.
- OSTI ID:
- 6872195
- Report Number(s):
- CONF-900143--
- Journal Information:
- IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) Vol. 37:2; ISSN 0018-9499; ISSN IETNA
- Country of Publication:
- United States
- Language:
- English
Similar Records
A compact position-sensitive detector for PET
High resolution block detectors for PET
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
550601 -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
654001 -- Radiation & Shielding Physics-- Radiation Physics
Shielding Calculations & Experiments
655002 -- Medical Physics-- Radiation Source Calibration & Standardization
73 NUCLEAR PHYSICS AND RADIATION PHYSICS
COMPUTERIZED TOMOGRAPHY
DESIGN
DIAGNOSTIC TECHNIQUES
DIGITIZERS
ELECTRONIC CIRCUITS
EMISSION COMPUTED TOMOGRAPHY
IMAGE PROCESSING
IMAGE SCANNERS
MEASURING INSTRUMENTS
MONTE CARLO METHOD
PERFORMANCE
PHOTOMULTIPLIERS
PHOTOTUBES
POSITION SENSITIVE DETECTORS
POSITRON COMPUTED TOMOGRAPHY
PROCESSING
PULSE CIRCUITS
RADIATION DETECTORS
SHIELDING
SIGNAL CONDITIONERS
SIMULATION
TECHNOLOGY ASSESSMENT
TOMOGRAPHY
TWO-DIMENSIONAL CALCULATIONS