Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

High resolution block detectors for PET

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6606784
; ; ;  [1]
  1. Hamamatsu Photonics K.K., 1126-1 Ichino-cho, Hamamatsu 435 (JP)

New high resolution block detectors were developed for positron emission tomographs (PET), in which BGO scintillators are coupled to a position-sensitive photomultiplier tube (PMT). The block detectors using two different structure of BGO scintillators were constructed and evaluated: one is 1.7 mm wide BGO array which provides the coincidence detector resolution less than 2 mm fwhm, and the other is a BGO block having comb-shape slits in it which has a potential capability of improving the resolution uniformity of a PET system because of its ability to detect the depth-of-interaction of {gamma}-rays in the scintillator. They are applied to a PET system for animal studies under development.

OSTI ID:
6606784
Report Number(s):
CONF-900143--
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) Vol. 37:2; ISSN 0018-9499; ISSN IETNA
Country of Publication:
United States
Language:
English