Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Development of a high resolution PET

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6867129
; ; ; ; ; ; ; ; ; ;  [1]; ; ; ; ;  [2]
  1. Hamamatsu Photonic K.K., 1126-1, Ichino-cho, Hamamatsu 435 (JP)
  2. National Inst. of Radiological Sciences, Chiba (Japan)

A high resolution positron emission tomograph (PET) for brain studies has been developed, which consists of 5 detector rings (240 BGO's/ring). New multi-segment photomultiplier tubes (PMT) were adopted to the system with 5 mm wide BGO's. The system is designed to examine a patient sitting or lying down on a chair/bed couch. The functions of PMT auto gain control and real time image display are implemented in the system. The physical performance of the system was evaluated: the spatial resolution is 3.5 mm in the transaxial plane and 5.7 mm in the axial direction, and the total system sensitivity is 109 kcps/{mu}Ci/ml for a 20 cm dia. uniform phantom with a pulse height threshold of 350 keV.

OSTI ID:
6867129
Report Number(s):
CONF-900143--
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) Vol. 37:2; ISSN 0018-9499; ISSN IETNA
Country of Publication:
United States
Language:
English