Basic imaging performance characteristics of a variable field of view PET camera using quadrant sharing detector design

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.775568· OSTI ID:684497
; ; ; ; ; ;  [1];  [2]
  1. Univ. of Texas M.D. Anderson Cancer Center, Houston, TX (United States)
  2. Univ. of Texas Health Science Center, Houston, TX (United States). Cyclotron Facility

The basic imaging performance of the prototype PET camera (MDAPET) developed in the laboratory was measured. The MDAPET was developed to test the engineering feasibility of two design concepts: (a) the photomultiplier-quadrant-sharing detector design and (b) a lower cost variable detector ring which forms a densely-packed smaller ring for imaging breast/brain/animals, and a less-packed ring for whole-body imaging. The basic imaging performance was measured for the 32 and 55 cm FOV with tests similar to those described in the SNM/NEMA performance standard. For the 32 cm FOV, the transaxial image resolutions at 0, 5, 10 cm were found to be 2.8, 3.1, 4.0 mm, respectively, while for the 55 cm FOV, the image resolutions were 2.9, 3.1, 3.6 mm. The axial resolution ranged from 2.5 to 4.4 mm as a function of detector-ring position. For a 21.5-cm uniform phantom, the coincidence-detection sensitivity was 82.3 kcps/{micro}Ci/cc for the whole body configuration. The scatter fractions were 19% and 30% for the whole-body mode and the brain/breast mode, respectively. The smaller mode has 2.4 times higher noise-equivalent-sensitivity than the larger mode at 1 {micro}Ci/cc. High quality, artifact-free brain-phantom images were obtained in both modes.

OSTI ID:
684497
Report Number(s):
CONF-981110--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 3Pt2 Vol. 46; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

The ECAT EXACT HR: Performance of a new high resolution positron scanner
Journal Article · Fri Dec 31 23:00:00 EST 1993 · Journal of Computer Assisted Tomography · OSTI ID:529510

A high resolution animal PET scanner using compact PS-PMT detectors
Conference · Mon Dec 30 23:00:00 EST 1996 · OSTI ID:513204

Development of a high resolution PET
Conference · Sat Mar 31 23:00:00 EST 1990 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) · OSTI ID:6867129