Effects of temperature on the response time of thermocouples and resistance thermometers

Conference ·
OSTI ID:6804827
The response times of sheathed thermocouples and platinum resistance thermometers both bare and in thermowells were measured at temperatures up to 650/sup 0/C using plunge tests and an in situ loop-current step-response (LCSR) measurement technique. Both methods of measurements agreed over a wide range of sensor sizes and response times, verifying the LCSR method. The response times were temperature dependent. The magnitude and direction of the temperature-dependence of response time varies from sensor to sensor. The response time of the sensor at the operating conditions must be measured when the sensor is at operating temperature.