Measurement of the transient response of thermocouples and resistance thermometers using an in situ method. [LMFBR]

Technical Report ·
DOI:https://doi.org/10.2172/7221079· OSTI ID:7221079
A loop-current step-response (LCSR) method was developed to measure, in situ, the transient response of temperature sensors, such as thermocouples and resistance thermometers. In this method, a sensor is heated with an electric current, and the time dependence of cooling is analyzed when the current is turned off. The cooling analysis is mathematically transformed into an equation to predict the transient response of the sensor to a change of external temperature. The method was verified for a limited class of sensors by comparing the predicted transient response with that measured after plunging the sensors into hot water. The transient responses of sheathed, insulated junction, Chromel/Alumel thermocouples of various diameters were measured with the sensing junction in flowing sodium at temperatures from 180 to 600/sup 0/C.