Measurement of cyclosporine concentrations in whole blood: HPLC and radioimmunoassay with a specific monoclonal antibody and /sup 3/H- or /sup 125/I-labeled ligand compared

Journal Article · · Clin. Chem. (Winston-Salem, N.C.); (United States)
OSTI ID:6500356
We compared cyclosporine concentrations in whole blood as measured by HPLC and by RIA with a monoclonal antibody specific for cyclosporine with /sup 3/H- or /sup 125/I-labeled cyclosporine ligand. The /sup 3/H-RIA kit slightly underestimated cyclosporine concentrations (greater than 600 micrograms/L) in comparison with HPLC. Over a wide range of concentrations, cyclosporine measured with the /sup 125/I-RIA kit correlated well with HPLC (slope = 0.99, n = 301, r = 0.98), observed for samples from recipients of kidney, heart, or liver allografts (respective slopes: 1.01, 0.93, and 1.00). The /sup 125/I-RIA standard curve was linear to 1000 micrograms of cyclosporine per liter. Inter- and intra-assay CVs for /sup 125/I-RIA measurements of cyclosporine were less than or equal to 7%. Evidently, the /sup 125/I-RIA kit involving a monoclonal antibody specific for cyclosporine is equivalent to the HPLC assay and can replace it for therapeutic drug monitoring of cyclosporine therapy.