Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Portable beta spectrometer for tissue dose measurement

Conference ·
OSTI ID:6399464
A technique for measuring the D(.07) and D(10) dose rates in mixed beta/gamma radiation fields through the use of a tissue equivalent detector and portable multichannel analyzer has been developed and field tested. Two energy spectra are collected for each measurement, an open window spectrum representing the tissue response to betas plus photons, and a second filtered spectrum representing the photon response. A close approximation to the net beta spectrum is obtained by spectrum stripping, and the D(.07) dose rate is calculated by multiplying channel by channel by a beta energy deposition function. Computer plots of typical field spectra and dose rate calculations are presented. 3 refs., 17 figs.
Research Organization:
EG and G Idaho, Inc., Idaho Falls (USA)
DOE Contract Number:
AC07-76ID01570
OSTI ID:
6399464
Report Number(s):
EGG-M-04985; CONF-8510185-2; ON: DE86004106
Country of Publication:
United States
Language:
English