Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Pulse shape discrimination in a portable beta-gamma dose rate meter

Journal Article · · IEEE Trans. Nucl. Sci.; (United States)
A portable beta-gamma dose rate instrument capable of tissue-equivalent dose rate measurements in mixed beta-gamma fields has been developed and tested. The detector is a 5 mg/cm/sup 2/-thick plastic scintillator with a 4 mg/cm/sup 2/-window and a 1 cm-thick lucite light pipe to achieve the backscatter geometry required for this application. Pulse shape discrimination techniques are used to minimize contributions from Cerenkov events within the lucite light pipe, photomultiplier tube glass, and direct interaction within the photomultiplier tube dynode structure. The response of the prototype instrument has been demonstrated to be independent of beta end-point energy to a precision of + or - 5% from 150 keV to 2.9 MeV and 1 mRem/h to 200 Rem/h. Response to gamma radiation in the energy range >400 keV is also accurate, but low energy photon response (<100 keV) is considerably lower than expected.
Research Organization:
EG and G Idaho, R.L. Dickson, DOE/RESL Idaho National Engineering Laboratory (INEL), P.O. Box 1625, Idaho Falls, Idaho 83415
DOE Contract Number:
AC07-76ID01570
OSTI ID:
5913405
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Journal Name: IEEE Trans. Nucl. Sci.; (United States) Journal Issue: 1 Vol. 30:1; ISSN IETNA
Country of Publication:
United States
Language:
English