Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A SCINTILLATION ALPHA-BETA-GAMMA FIXED-FILTER COUNTER

Technical Report ·
OSTI ID:4314549
BS>A scintillation alpha-beta-gamma filter counter has been developed which simultaneously monitors filters for alpha ajn beta-gamma contamination with none of the faults of the present equipment. The instrument has two lucite light pipes, two RCA 6655 photomultiplier tubes, two scalnag units, and a light-tight box with a roller bearing mounted slide for transporting the filter. The average center geometry for the alpha probe is 20 to 25 per cent wlth a background of approximately 2 c/m. The average center efficiency for the beta-gamma probe is 0.8 to 1.0 per cent for a 600-kev gamma source and 20 to 25 per cent for a 500- kev beta source. (auth)
Research Organization:
General Electric Co. Hanford Atomic Products Operation, Richland, Wash.
DOE Contract Number:
W-31-109-ENG-52
NSA Number:
NSA-12-012573
OSTI ID:
4314549
Report Number(s):
HW-54947(Rev.)
Country of Publication:
United States
Language:
English

Similar Records

AN ALPHA, BETA, GAMMA HAND AND SHOE COUNTER-MODEL II
Technical Report · Sun Jun 01 00:00:00 EDT 1958 · OSTI ID:4314550

DESIGN AND PERFORMANCE OF RECTANGULAR LARGE-AREA SCINTILLATION COUNTERS
Journal Article · Sat Nov 30 23:00:00 EST 1963 · Nucl. Instr. Methods · OSTI ID:4039219

Pulse shape discrimination in a portable beta-gamma dose rate meter
Journal Article · Mon Jan 31 23:00:00 EST 1983 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:5913405