Field tests of a portable tissue equivalent survey meter for monitoring mixed beta/gamma radiation fields
Technical Report
·
OSTI ID:5746438
A portable radiation survey meter that provides a tissue equivalent response to photons and beta particles has been designed and field tested. The detector is a very thin plastic scintillator that closely simulates the actual geometry and scattering properties of the relevant skin tissues. The meter reads out the D(0.07) dose rate directly, and indicates the tissue dose rates at other depths with the use of tissue equivalent filters of appropriate thicknesses. Data are presented which compare the D(0.07) and D(10) dose rates recorded by the Tissue Equivalent (TE) survey meter with dose rates recorded by two commercial ion chamber meters for a number of laboratory and field sources. Most commercial ion chamber meters fail to respond adequately to the extreme off-axis beta particles from extended beta sources, and hence require the application of large beta correction factors to change the instrument reading to the true D(0.07) dose rate. The tissue equivalent survey meter exhibits an angular response to beta particles that is very similar to the angular response of an extrapolation chamber. Consequently, there is close agreement between the TE meter and extrapolation chamber readings for a wide variety of beta and mixed beta-gamma rdiation fields. D(0.07), D(3), and D(10) dose rates, measured with the INEL TE meter at a number of typical work stations, are presented.
- Research Organization:
- EG and G Idaho, Inc., Idaho Falls (USA)
- DOE Contract Number:
- AC07-76ID01570
- OSTI ID:
- 5746438
- Report Number(s):
- NUREG/CR-4553; EGG-2448; ON: TI86013018
- Country of Publication:
- United States
- Language:
- English
Similar Records
BETA FIELD DOSE RATE DETERMINATION FOR ANGULAR RESPONSE WITH THE SOUP BOWL DOSE RATE METER
Portable beta spectrometer for tissue dose measurement
``Hot particle`` intercomparison dosimetry
Technical Report
·
Thu Sep 12 00:00:00 EDT 1957
·
OSTI ID:4816387
Portable beta spectrometer for tissue dose measurement
Conference
·
Mon Dec 31 23:00:00 EST 1984
·
OSTI ID:6399464
``Hot particle`` intercomparison dosimetry
Conference
·
Sat Jun 01 00:00:00 EDT 1996
·
OSTI ID:277001
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BETA DETECTION
CHARGED PARTICLE DETECTION
DETECTION
DOSE RATES
EQUIPMENT
FIELD TESTS
GAMMA DETECTION
MEASURING INSTRUMENTS
MONITORS
PHOSPHORS
PLASTIC SCINTILLATORS
PORTABLE EQUIPMENT
RADIATION DETECTION
RADIATION DETECTORS
RADIATION MONITORS
SURVEY MONITORS
TESTING
TISSUE-EQUIVALENT DETECTORS
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BETA DETECTION
CHARGED PARTICLE DETECTION
DETECTION
DOSE RATES
EQUIPMENT
FIELD TESTS
GAMMA DETECTION
MEASURING INSTRUMENTS
MONITORS
PHOSPHORS
PLASTIC SCINTILLATORS
PORTABLE EQUIPMENT
RADIATION DETECTION
RADIATION DETECTORS
RADIATION MONITORS
SURVEY MONITORS
TESTING
TISSUE-EQUIVALENT DETECTORS