Graphite furnace atomic absorption spectrometry with graphite cloth ribbon for selenium determination

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00268a059· OSTI ID:6054493

In this study, the sensitivity for selenium is reported to depend heavily on the nature of the graphite surface; sensitivities obtained by using a nonpyrolytic graphite tube are higher than those obtained by using a pyrolytic graphite tube; and the volatilization loss of selenium in the charring process is less critical for the nonpyrolytic graphite tube. Furthermore, it is reported that the use of a graphite cloth ribbon placed inside the graphite tube causes a large contact area with analytes and greatly enhances the sensitivity in selenium atomic absorption signals.

Research Organization:
Hiroshima Univ., Japan
OSTI ID:
6054493
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 56:4; ISSN ANCHA
Country of Publication:
United States
Language:
English