Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Chloride interferences in graphite furnace atomic absorption spectrometry

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00266a011· OSTI ID:5965588
The premature loss of analyte as a volatile halide is shown to be an important interference in some situations. When this char step interference is avoided, an interference in the atomization step can arise when large amounts of halide are present during that step. This atomization interference results from the binding of the analyte as a vapor phase metal halide, thereby preventing some portion of the analyte from absorbing atomic radiation. This vapor phase interference can be circumvented in several ways. The most convenient is the use of high concentrations of HNO/sub 3/ or H/sub 2/SO/sub 4/ which drives off more of the chloride as HCl during the dry and char steps. As much CaCl/sub 2/ in solution as 1 mg reduced the Mn signal in the presence of HNO/sub 3/ and Mg(NO/sub 3/)/sub 2/ by only 15%. Analogous experience was found with Tl. We believe the experience is quite general. The halide which interferes in the vapor phase is retained from the char step by adsorption or intercalation on the graphite tube. Thus we can expect still further improvements in the handling of large amounts of halide matrix when glassy carbon or solid pyroltic tubes become available.
Research Organization:
Perkin-Elmer Corp., Norwalk, CT
OSTI ID:
5965588
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 56:2; ISSN ANCHA
Country of Publication:
United States
Language:
English