Evaluation of commercially available TLD readers
Technical Report
·
OSTI ID:5893402
The performance of two commercially available thermoluminescence bulb dosimetry systems, the Victoreen 2810 Ultra Sensitive TLD Reader and the Harshaw Model 2000P Bulb Dosimetry System, was investigated. Of primary interest was the reliability of the instruments at an exposure range typically encountered in environmental radiation when monitoring on a monthly basis, i.e., approx. 10 mR. Uniformity, linearity, and reproducibility test results are reported. The two systems were found to perform comparably well as bulb readers. Evaluation of the dual capability of the Victoreen 2810 as a chip reader is also described.
- Research Organization:
- USDOE Environmental Measurements Lab., New York
- OSTI ID:
- 5893402
- Report Number(s):
- DOE/EML-396; ON: DE82002447
- Country of Publication:
- United States
- Language:
- English
Similar Records
Preirradiation sensitivity loss of reader-annealed LiF TLDs
Comparison of environmental TLD (thermoluminescent dosimeter) results obtained using glow curve deconvolution and region of interest analysis
Thermoluminescent dosimeter (TLD). IV. Popular type TLD reader
Conference
·
Wed Dec 31 23:00:00 EST 1986
· Trans. Am. Nucl. Soc.; (United States)
·
OSTI ID:5351135
Comparison of environmental TLD (thermoluminescent dosimeter) results obtained using glow curve deconvolution and region of interest analysis
Conference
·
Wed Dec 31 23:00:00 EST 1986
·
OSTI ID:5665482
Thermoluminescent dosimeter (TLD). IV. Popular type TLD reader
Journal Article
·
Mon Oct 01 00:00:00 EDT 1973
· Nat. Tech. Rep. (Matsushita Elec. Ind. Co., Osaka), v. 19, no. 5, pp. 479-487
·
OSTI ID:4352442