Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Preirradiation sensitivity loss of reader-annealed LiF TLDs

Conference · · Trans. Am. Nucl. Soc.; (United States)
OSTI ID:5351135
Yankee Atomic Electric Company (YAEC) maintains a thermoluminescent dosimeter (TLD) dosimetry service using a Harshaw TLD badge system consisting of two TLD cards in a plastic case and automatic 2271 readers. Beginning in 1986, a standardized on-site quality assurance testing program was instituted for the facilities provided with YAEC personnel dosimetry. Results obtained during the first half of 1986 indicated a low bias in test TLD results on the order of 10 to 15%. A subsequent investigation attributed this bias to a previously unidentified loss in the sensitivity of reader-annealed LiF TLDs with storage time prior to irradiation. The discovery of this effect, based on in-plant quality assurance testing, demonstrates the usefulness of even relatively simple quality assurance tests by users themselves, regardless of the apparent thoroughness of other testing.
OSTI ID:
5351135
Report Number(s):
CONF-870837-
Conference Information:
Journal Name: Trans. Am. Nucl. Soc.; (United States) Journal Volume: 54
Country of Publication:
United States
Language:
English