Radioimmunoassay and chemical ionization/mass spectrometry compared for plasma cortisol determination

Journal Article · · Clin. Chem. (Winston-Salem, N.C.); (United States)
OSTI ID:5639362

A method is described for determination of cortisol in plasma and urine, based on chemical ionization/mass spectrometry with deuterium-labeled cortisol as the internal standard. The within-run precision (CV) was 2.5-5.7%, the between-run precision 4.6%. Results by this method were compared with those by a radioimmunological method (RIANEN Cortisol, New England Nuclear) for 395 plasma samples. The latter method gave significantly higher (approx. 25%) cortisol values.

OSTI ID:
5639362
Journal Information:
Clin. Chem. (Winston-Salem, N.C.); (United States), Journal Name: Clin. Chem. (Winston-Salem, N.C.); (United States) Vol. 28:1; ISSN CLCHA
Country of Publication:
United States
Language:
English