Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Scintillation hodoscope with working area of 50 x 50 cm based on hodoscopic photomultipliers

Journal Article · · Instrum. Exp. Tech. (Engl. Transl.); (United States)
OSTI ID:5213834
The choice of optimum designs for the sensitive elements of large hodoscopes based on hodoscopic photomultipliers is examined. The results of numerical calculations are confirmed by measurement results. The measured space resolution of one of the scintillation-hodoscope designs (with two hodoscopic photomultipliers) with a sensitive volume of 50 x 50 x 2 mm is less than or equal to +3 mm.