First use of a mercuric iodide (HgI/sub 2/) energy dispersive x-ray detector in a scanning electron microscope

Journal Article · · Scanning Electron Microsc.; (United States)
OSTI ID:5130856

A room temperature mercuric iodide x-ray detector system with a Peltier cooled field effect transistor and pulsed optical feedback preamplifier has been constructed and tested in a scanning electron microscope. The initial system achieved an energy resolution of 220 eV (FWHM) for Mg-K line at 1.25 keV and 360 eV for Mn-K..cap alpha.. at 5.9 keV. This resolution is adequate for many applications. Improved energy resolution and low energy sensitivity should be readily achievable through simple changes to the not-optimized system tested.