"Total Body Counter" of the C.E.N. at Mol. Calibration and Observations of 1500 Measurements; "TOTAL BODY COUNTER" DU C.E.N. A MOL ETALONNAGE ET OBSERVATIONS SUR 1.500 MESURES
Technical Report
·
OSTI ID:4842965
A device for the detection of body burden radioactivity is reported. The measurement is done in a steel shielded room by gamma spectrometry with a NaI crystal 8-in. diameter x 4-in. height; the subject is placed in a standard position on a chair below the crystal. The method was used for 1500 measurements in routine work or after accidents, and the observed contaminations are discussed. The measurements have shown a noticeable decrease since 2 years of the body burden of Cs/sup 137/ caused by fallout and the amount of natural potassium in the body was measured. (auth)