Low Voltage He3-Filled Proportional Counter for Efficient Detection of Thermal and Epithermal Neutrons

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1717994· OSTI ID:4805473
Neutron counters using He3 gas at absolute pressures up to 10 atm were constructed and tested. The relative detection efficiency for thermal, epicadmium, and fast neutrons was measured as a function of pressure from 1 to 10 atm and compared with similar data from a boron trifluoride counter of the same size filled to a pressure of 65 cm of B10F3 gas. The ratio He3/ BF3 increases approximately linearly with pressure for epicadmium and fast neutrons, reaehing values of 26.6 and 21.6, respectively, at 10 atm He3 pressure. This ratio shows a saturation effect for pressure. The gamma-ray response was found to be low with proper pulse shaping. High voltage plateau curves for 3- and 7atm counters are given and indicate operating regions at least 100 v wide at a counter high voltage of about 1200 v.
Research Organization:
Socony Mobil Oil Co., Inc., Dallas, Tex.
Sponsoring Organization:
USDOE
NSA Number:
NSA-16-029168
OSTI ID:
4805473
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 33; ISSN 0034-6748; ISSN RSINAK
Publisher:
American Institute of Physics (AIP)
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

Very Large Boron Trifluoride Proportional Counters
Journal Article · Mon Dec 31 23:00:00 EST 1962 · Review of Scientific Instruments · OSTI ID:4670758

SENSITIVE LOW VOLTAGE PROPORTIONAL COUNTER FOR NEUTRON DETECTION CONTAINING He$sup 3$ AT SUPERATMOSPHERIC PRESSURE
Patent · Tue Aug 27 00:00:00 EDT 1963 · OSTI ID:4129758

SENSITIVE LOW VOLTAGE PROPORTIONAL COUNTER
Patent · Wed Jul 11 00:00:00 EDT 1962 · OSTI ID:4792165