PRACTICAL ASPECTS OF FERROUS SULPHATE DOSIMETRY
Journal Article
·
· Physics in Medicine and Biology (England)
Practical details in accurate ferrous sulfate dosimetry are discussed, including a new method for estimating the extinction coefficient of Fe/sup 3+/ ion. Errors in the ferrous sulfate G-value amounting to several per cent can arise due to the presence of impurities in the analytical reagent grade sulfuric acid normally used to make up the dosimeter solution. Simple methods of removing this difficulty are described. (auth)
- Research Organization:
- Christie Hospital and Holt Radium Inst., Manchester, Eng.
- NSA Number:
- NSA-17-020078
- OSTI ID:
- 4732116
- Journal Information:
- Physics in Medicine and Biology (England), Journal Name: Physics in Medicine and Biology (England) Vol. Vol: 8; ISSN PHMBA
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
A PITFALL TO AVOID IN FERROUS SULPHATE DOSIMETRY
HIGH DOSE FERROUS DOSIMETRY FOR Co-60 $gamma$-RAYS
THE G VALUE AND THE REPRODUCIBILITY OF THE FERROUS-FERRIC DOSIMETER
Journal Article
·
Thu Jun 01 00:00:00 EDT 1961
· Brit. J. Radiol.
·
OSTI ID:4046122
HIGH DOSE FERROUS DOSIMETRY FOR Co-60 $gamma$-RAYS
Journal Article
·
Mon Dec 31 23:00:00 EST 1962
· Proc. Japan Conf. Radioisotopes
·
OSTI ID:4690903
THE G VALUE AND THE REPRODUCIBILITY OF THE FERROUS-FERRIC DOSIMETER
Technical Report
·
Tue Oct 01 00:00:00 EDT 1963
·
OSTI ID:4110684