Study on Photomultipliers with Optical Impulses; UNTERSUCHUNGEN AN PHOTOMULTIPLIERN MIT OPTISCHEN IMPULSEN

Technical Report ·
OSTI ID:4671333
A test-apparatus was developed for measuring the characteristics of photomultipliers in the nanosecond range. With periodical light-pulses (1/2 nsec rise- and falltime, 2 nsec pulse width at half maximum) the spread of transit time with respect to the illumination point on the photocathode, the total transit time, the risetime of the output signal and the gain can be measured as a function of parameters like high voltage or electrode-voltages. The output pulses are displayed on a Tektronix Sampling oscilloscope, Type 661. Different photomultipliers can be tested, and it is possible to find out the most qualified photomultiplier for a specified application. As an example the measurements made with a Philips 56AVP are described. (auth)
Research Organization:
Deutsches Elektronen-Synchrotron, Hamburg
NSA Number:
NSA-17-032335
OSTI ID:
4671333
Report Number(s):
DESY-A2.99
Country of Publication:
Germany
Language:
English

Similar Records

TESTS OF NEW FAST PHOTOMULTIPLIERS
Journal Article · Wed Aug 01 00:00:00 EDT 1962 · Bull. Inform. Sci. Tech. (Paris) · OSTI ID:4730887

Measurement of spread of transmission times of single-photoelectron signals through FEU-49B photomultiplier
Journal Article · Mon Aug 01 00:00:00 EDT 1988 · Instrum. Exp. Tech. (Engl. Transl.); (United States) · OSTI ID:6284652

Performance of studies of Hamamatsu R647-01 photomultiplier
Conference · Tue Jan 31 23:00:00 EST 1984 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:6920060