Performance of studies of Hamamatsu R647-01 photomultiplier
Conference
·
· IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:6920060
Characteristics of the Hamamatsu R647-01 13 mm-diameter photomultiplier have been measured. Some typical photomultiplier characteristics-such as gain, dark current, transit and rise times-are compared with data provided by the manufacturer. Other photomultiplier characteristics, generally not available from the manufacturer, such as the single photoelectron time spread and pulse response for the full photocathode illumination were measured and are discussed. In particular the optimization of voltage divider for photomultiplier was investigated for timing applications in positron emission tomography systems.
- Research Organization:
- Lawrence Berkeley Laboratory, University of California, Berkeley, CA 94720
- OSTI ID:
- 6920060
- Report Number(s):
- CONF-831015-
- Conference Information:
- Journal Name: IEEE Trans. Nucl. Sci.; (United States) Journal Volume: 31:1
- Country of Publication:
- United States
- Language:
- English
Similar Records
Evaluation of Hamamatsu R1635 photomultiplier
Time resolution performance studies of Hamamatsu R1564U microchannel plate photomultiplier
Evaluation of Hamamatsu R1635 photomultiplier
Technical Report
·
Sun May 01 00:00:00 EDT 1983
·
OSTI ID:5665042
Time resolution performance studies of Hamamatsu R1564U microchannel plate photomultiplier
Conference
·
Sat Jan 31 23:00:00 EST 1987
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:7069549
Evaluation of Hamamatsu R1635 photomultiplier
Journal Article
·
Sun Jul 01 00:00:00 EDT 1984
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:6888819
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFICATION
CATHODES
COMPUTERIZED TOMOGRAPHY
COUNTING TECHNIQUES
CURRENTS
DIAGNOSTIC TECHNIQUES
ELECTRIC CURRENTS
ELECTRICAL EQUIPMENT
ELECTRODES
EMISSION COMPUTED TOMOGRAPHY
EQUIPMENT
GAIN
ILLUMINANCE
OPTIMIZATION
PERFORMANCE
PHOTOCATHODES
PHOTOCURRENTS
PHOTOELECTRON COUNTING
PHOTOMULTIPLIERS
PHOTORESISTORS
PHOTOSENSITIVITY
PHOTOTUBES
POSITRON COMPUTED TOMOGRAPHY
PULSE RISE TIME
RESISTORS
RESOLUTION
SENSITIVITY
TECHNOLOGY UTILIZATION
TIME MEASUREMENT
TIME RESOLUTION
TIMING PROPERTIES
TOMOGRAPHY
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFICATION
CATHODES
COMPUTERIZED TOMOGRAPHY
COUNTING TECHNIQUES
CURRENTS
DIAGNOSTIC TECHNIQUES
ELECTRIC CURRENTS
ELECTRICAL EQUIPMENT
ELECTRODES
EMISSION COMPUTED TOMOGRAPHY
EQUIPMENT
GAIN
ILLUMINANCE
OPTIMIZATION
PERFORMANCE
PHOTOCATHODES
PHOTOCURRENTS
PHOTOELECTRON COUNTING
PHOTOMULTIPLIERS
PHOTORESISTORS
PHOTOSENSITIVITY
PHOTOTUBES
POSITRON COMPUTED TOMOGRAPHY
PULSE RISE TIME
RESISTORS
RESOLUTION
SENSITIVITY
TECHNOLOGY UTILIZATION
TIME MEASUREMENT
TIME RESOLUTION
TIMING PROPERTIES
TOMOGRAPHY