Evaluation of Hamamatsu R1635 photomultiplier
Technical Report
·
OSTI ID:5665042
Characteristics have been measured of the Hamamatsu R1635 10 mm-diameter photomultiplier. Some typical photomultiplier characteristics - such as gain, dark current, transit and rise times - are compared with data provided by the manufacturer. Photomultiplier characteristics, generally not available from the manufacturer, such as the single photoelectron and pulse response time spread for full photocathode illumination were measured and are discussed.
- Research Organization:
- Lawrence Berkeley Lab., CA (USA)
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 5665042
- Report Number(s):
- LBL-16017; ON: DE84001252
- Country of Publication:
- United States
- Language:
- English
Similar Records
Evaluation of Hamamatsu R1635 photomultiplier
Evaluation of Hamamatsu R1635 photomultiplier
Performance of studies of Hamamatsu R647-01 photomultiplier
Journal Article
·
Sun Jul 01 00:00:00 EDT 1984
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:6213041
Evaluation of Hamamatsu R1635 photomultiplier
Journal Article
·
Sun Jul 01 00:00:00 EDT 1984
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:6888819
Performance of studies of Hamamatsu R647-01 photomultiplier
Conference
·
Tue Jan 31 23:00:00 EST 1984
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:6920060
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFICATION
CATHODES
CURRENTS
EFFICIENCY
ELECTRODES
EVALUATION
GAIN
PHOTOCATHODES
PHOTOMULTIPLIERS
PHOTOTUBES
PULSE RISE TIME
QUANTUM EFFICIENCY
RESOLUTION
TIME RESOLUTION
TIMING PROPERTIES
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFICATION
CATHODES
CURRENTS
EFFICIENCY
ELECTRODES
EVALUATION
GAIN
PHOTOCATHODES
PHOTOMULTIPLIERS
PHOTOTUBES
PULSE RISE TIME
QUANTUM EFFICIENCY
RESOLUTION
TIME RESOLUTION
TIMING PROPERTIES