THE ENERGY RESPONSE OF A VICTOREEN CONDENSER IONIZATION CHAMBER

Journal Article · · Phys. in Med. Biol.
The energy response of a 25-r Victoreen chamber has been measured over the range of photon energies from 40 kev to 1.25 Mev. The difference in response frequently observed between the conventional x-ray region and the gamma-ray region is confirmed. Theoretical response curves for chambers composed of combinations of graphite and heavier elements are compared with the Victoreen curve. (auth)
Research Organization:
National Research Council, Ottawa
NSA Number:
NSA-13-009169
OSTI ID:
4277655
Journal Information:
Phys. in Med. Biol., Journal Name: Phys. in Med. Biol. Vol. Vol: 3
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

A THEORY FOR HETEROGENEOUS ION CHAMBERS AND NEUTRONS RESPONSE OF VICTOREEN DOSIMETERS
Journal Article · Tue Oct 01 00:00:00 EDT 1957 · Radiation Research · OSTI ID:4324213

ENERGY RESPONSE OF LANDSVERK IONIZATION CHAMBERS TO MONOENERGETIC GAMMA RAYS
Technical Report · Tue May 01 00:00:00 EDT 1962 · OSTI ID:4764125

Measurement of Stopping-Power Ratio for Cavity Ionization Chambers
Journal Article · Fri Nov 10 23:00:00 EST 1961 · Nature (London) · OSTI ID:4811585