Measurement of Stopping-Power Ratio for Cavity Ionization Chambers
An absolute measurement of the stopping-power ratio is made by varying the pressure of different gases introduced into a single ionization chamber. The method is shown graphically by showing the variation in the ionization/g of gas with the gas pressure for hydrogen, air, and argon tn a graphite ionization chamber irradiated by 1.25-Mev photons (Co60 gamma rays).
- Research Organization:
- Hammersmith Hospital, London
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-16-004355
- OSTI ID:
- 4811585
- Journal Information:
- Nature (London), Journal Name: Nature (London) Journal Issue: 4802 Vol. 192; ISSN 0028-0836
- Publisher:
- Nature Publishing Group
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
                                
                                
                                    
                                        
                                        Investigation of the initial and volume recombination losses in gamma versatile cylindrical ionization chamber VGIC developed for gamma ray dosimetry
                                        
IONIZATION CHAMBER
THE STOPPING-POWER CORRECTION FOR GRAPHITE CAVITY-CHAMBERS USED WITH 2 mV X- RAYS
                        
                                            Conference
                                            ·
                                            Wed Jul 01 00:00:00 EDT 2015
                                            
                                            ·
                                            OSTI ID:22531471
                                        
                                        
                                        
                                    
                                
                                    
                                        IONIZATION CHAMBER
                                            Patent
                                            ·
                                            Wed Jul 11 00:00:00 EDT 1962
                                            
                                            ·
                                            OSTI ID:4817572
                                        
                                        
                                        
                                    
                                
                                    
                                        THE STOPPING-POWER CORRECTION FOR GRAPHITE CAVITY-CHAMBERS USED WITH 2 mV X- RAYS
                                            Journal Article
                                            ·
                                            Wed Jul 01 00:00:00 EDT 1959
                                            · Phys. in Med. Biol.
                                            ·
                                            OSTI ID:4227680