Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

CONICAL PLASTIC SCINTILLATORS SHOW TOTAL GAMMA ABSORPTION

Journal Article · · Nucleonics (U.S.) Ceased publication
OSTI ID:4151534
Total photon absorption, at least with low energy gamma and x rays, was observed in plastic scintillators. The best results were obtained with lead- loaded, conically-shaped scintillators surrounded by aluminuru oxide as a reflector. Low energy photons gave better absorption curves even poorer resolution than the higher energy photons. The best stintillation size for total absorption was found to be a function of energy. Surrounding the scintillator with aluminuru oxide instead of aluminum foil increased the pulse height slightly. The light output was increased by changing the shape from a cylinder to a truncated cone. By adding 2 wt.% lead to the plastic, the absorption of 123 kev gamma rays was increased by a factor of aporoximately 16, but reduced light output made resolution worse. (M.C.G.)
Research Organization:
Veterans Administration Hospital, Boston; and Boston Univ. School of Medicine
NSA Number:
NSA-14-023051
OSTI ID:
4151534
Journal Information:
Nucleonics (U.S.) Ceased publication, Journal Name: Nucleonics (U.S.) Ceased publication Vol. Vol: 18, No. 9; ISSN NUCLA
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

Energy resolution experiments of conical organic scintillators and a comparison with Geant4 simulations
Journal Article · Tue May 15 20:00:00 EDT 2018 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1798640

Plastic Scintillators Light Yield Energy Calibration.
Journal Article · Mon Nov 30 23:00:00 EST 2015 · OSTI ID:1427195

Plastic scintillators relative gamma-ray light yield measurement
Journal Article · Tue Feb 28 19:00:00 EST 2017 · Review of Scientific Instruments · OSTI ID:1333614