Energy resolution experiments of conical organic scintillators and a comparison with Geant4 simulations
Journal Article
·
· Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Univ. of Michigan, Ann Arbor, MI (United States); University of Michigan
- Idaho National Lab. (INL), Idaho Falls, ID (United States)
- Univ. of Michigan, Ann Arbor, MI (United States)
An increase in light-collection efficiency (LCE) improves the energy resolution of scintillator-based detection systems. An improvement in energy resolution can benefit detector performance, for example by lowering the measurement threshold and achieving greater accuracy in light-output calibration. This work shows that LCE can be increased by modifying the scintillator shape to reduce optical-photon reflections, thereby decreasing transmission and absorption likelihood at the reflector boundary. The energy resolution of four organic scintillators (EJ200) were compared: two cones and two right-circular cylinders, all with equal base diameter and height (50 mm). The sides of each shape had two surface conditions: one was polished and the other was ground. Each scintillator was coupled to the center of four photomultiplier tube (PMT) configurations of different diameters. The photocathode response of all PMTs was assessed as a function of position using a small cube (5 mm height) of EJ200. The worst configuration, a highly polished conical scintillator mated to a PMT of equal base diameter, produced a smeared energy spectrum. The cause of spectrum smearing is explored in detail. Results demonstrate that a ground cone had the greatest improvement in energy resolution over a ground cylinder by approximately 16.2% at 478 keVee, when using the largest diameter (127 mm) PMT. Furthermore, this result is attributed to the greater LCE of the cone, its ground surface, and the uniform photocathode response near center of the largest PMT. Optical-photon transport simulations in Geant4 of the cone and cylinder assuming a diffuse reflector and a uniform photocathode were compared to the best experimental configuration and agreed well. If a detector application requires excellent energy resolution above all other considerations, a ground cone on a large PMT is recommended over a cylinder.
- Research Organization:
- Univ. of Michigan, Ann Arbor, MI (United States)
- Sponsoring Organization:
- USDOE; USDOE National Nuclear Security Administration (NNSA), Office of Nonproliferation and Verification Research and Development
- Grant/Contract Number:
- AC07-05ID14517; NA0002534
- OSTI ID:
- 1798640
- Alternate ID(s):
- OSTI ID: 1542495
OSTI ID: 23064845
- Journal Information:
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Journal Name: Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment Vol. 898; ISSN 0168-9002
- Publisher:
- ElsevierCopyright Statement
- Country of Publication:
- United States
- Language:
- English
Time resolution measurements with an improved discriminator and conical scintillators
|
journal | June 1977 |
Development of organic scintillators
|
journal | June 1979 |
A study of light collection efficiency in scintillation detectors
|
journal | December 1984 |
Contribution of photocathode nonuniformity to energy resolution in NaI(Tl) scintillation detectors
|
journal | March 1990 |
Geant4—a simulation toolkit
|
journal | July 2003 |
Geometry and surface treatment dependence of the light collection from LSO crystals
|
journal | November 1999 |
Light collection optimization in scintillator-based gamma-ray spectrometers
|
journal | April 2008 |
Energy calibration of gamma spectra in plastic scintillators using Compton kinematics
|
journal | September 2008 |
Rattling nucleons: New developments in active interrogation of special nuclear material
|
journal | January 2012 |
Dual-particle imaging system based on simultaneous detection of photon and neutron collision events
|
journal | October 2014 |
Pulse shape discrimination performance of stilbene coupled to low-noise silicon photomultipliers
|
journal | September 2015 |
Digital pulse processing for NaI(Tl) detectors
|
journal | January 2016 |
Passive assay of plutonium metal plates using a fast-neutron multiplicity counter
|
journal | May 2017 |
Impact of geometry on light collection efficiency of scintillation detectors for cryogenic rare event searches
|
journal | October 2014 |
Measurement of Compton edge position in low-Z scintillators
|
journal | March 2010 |
Photocathode non-uniformity contribution to the energy resolution of scintillators
|
journal | February 2010 |
The development of large-area plastic gamma-ray spectrometers
|
conference | October 2008 |
Characterization and calibration of large-volume PVT detectors by Backscatter Gating
|
conference | October 2011 |
A study for the characterization of high QE photocathodes
|
conference | June 2007 |
Resolution and Cathode Uniformity in Scintillation Counters
|
journal | January 1967 |
Improvement of the Light Collection in Scintillation Detectors
|
journal | January 1979 |
Measuring Light Reflectance of BGO Crystal Surfaces
|
journal | October 2008 |
Similar Records
Position Sensitive Photomultiplier Tubes for scintillation imaging
Time Resolution Measurements of EJ-232Q With Single- and Dual-Sided Readouts
A simple method to improve the spatial uniformity of venetian-blind photomultiplier tubes
Conference
·
Fri Jan 31 23:00:00 EST 1986
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:6031865
Time Resolution Measurements of EJ-232Q With Single- and Dual-Sided Readouts
Journal Article
·
Sun Jul 19 20:00:00 EDT 2020
· IEEE Transactions on Nuclear Science
·
OSTI ID:1826752
A simple method to improve the spatial uniformity of venetian-blind photomultiplier tubes
Journal Article
·
Sat Jun 01 00:00:00 EDT 1996
· IEEE Transactions on Nuclear Science
·
OSTI ID:277665