Title: Energy resolution experiments of conical organic scintillators and a comparison with Geant4 simulations

Journal Article · · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
 [1];  [2];  [2];  [1];  [1];  [1]
  1. Univ. of Michigan, Ann Arbor, MI (United States)
  2. Idaho National Lab. (INL), Idaho Falls, ID (United States)

An increase in light-collection efficiency (LCE) improves the energy resolution of scintillator-based detection systems. An improvement in energy resolution can benefit detector performance, for example by lowering the measurement threshold and achieving greater accuracy in light-output calibration. This work shows that LCE can be increased by modifying the scintillator shape to reduce optical-photon reflections, thereby decreasing transmission and absorption likelihood at the reflector boundary. The energy resolution of four organic scintillators (EJ200) were compared: two cones and two right-circular cylinders, all with equal base diameter and height (50 mm). The sides of each shape had two surface conditions: one was polished and the other was ground. Each scintillator was coupled to the center of four photomultiplier tube (PMT) configurations of different diameters. The photocathode response of all PMTs was assessed as a function of position using a small cube (5 mm height) of EJ200. The worst configuration, a highly polished conical scintillator mated to a PMT of equal base diameter, produced a smeared energy spectrum. The cause of spectrum smearing is explored in detail. Results demonstrate that a ground cone had the greatest improvement in energy resolution over a ground cylinder by approximately 16.2% at 478 keVee, when using the largest diameter (127 mm) PMT. Furthermore, this result is attributed to the greater LCE of the cone, its ground surface, and the uniform photocathode response near center of the largest PMT. Optical-photon transport simulations in Geant4 of the cone and cylinder assuming a diffuse reflector and a uniform photocathode were compared to the best experimental configuration and agreed well. If a detector application requires excellent energy resolution above all other considerations, a ground cone on a large PMT is recommended over a cylinder.

Research Organization:
Univ. of Michigan, Ann Arbor, MI (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA), Office of Nonproliferation and Verification Research and Development; USDOE
Grant/Contract Number:
NA0002534; AC07-05ID14517
OSTI ID:
1798640
Alternate ID(s):
OSTI ID: 1542495
Journal Information:
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 898; ISSN 0168-9002
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 5 works
Citation information provided by
Web of Science

References (22)

Rattling nucleons: New developments in active interrogation of special nuclear material
  • Runkle, Robert C.; Chichester, David L.; Thompson, Scott J.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 663, Issue 1 https://doi.org/10.1016/j.nima.2011.09.052
journal January 2012
Time resolution measurements with an improved discriminator and conical scintillators journal June 1977
A study for the characterization of high QE photocathodes conference June 2007
Photocathode non-uniformity contribution to the energy resolution of scintillators journal February 2010
Energy calibration of gamma spectra in plastic scintillators using Compton kinematics
  • Siciliano, E. R.; Ely, J. H.; Kouzes, R. T.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 594, Issue 2 https://doi.org/10.1016/j.nima.2008.06.031
journal September 2008
Impact of geometry on light collection efficiency of scintillation detectors for cryogenic rare event searches
  • Danevich, F. A.; Kobychev, V. V.; Kobychev, R. V.
  • Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 336 https://doi.org/10.1016/j.nimb.2014.06.015
journal October 2014
Light collection optimization in scintillator-based gamma-ray spectrometers
  • Hull, G.; Du, S.; Niedermayr, T.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 588, Issue 3 https://doi.org/10.1016/j.nima.2008.01.092
journal April 2008
Geometry and surface treatment dependence of the light collection from LSO crystals
  • Huber, J. S.; Moses, W. W.; Andreaco, M. S.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 437, Issue 2-3 https://doi.org/10.1016/S0168-9002(99)00766-4
journal November 1999
Development of organic scintillators journal June 1979
The development of large-area plastic gamma-ray spectrometers
  • Burt, Christopher; Ramsden, David
  • 2008 IEEE Nuclear Science Symposium and Medical Imaging conference (2008 NSS/MIC), 2008 IEEE Nuclear Science Symposium Conference Record https://doi.org/10.1109/NSSMIC.2008.4774614
conference October 2008
Measuring Light Reflectance of BGO Crystal Surfaces journal October 2008
Passive assay of plutonium metal plates using a fast-neutron multiplicity counter
  • Di Fulvio, A.; Shin, T. H.; Jordan, T.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 855 https://doi.org/10.1016/j.nima.2017.02.082
journal May 2017
Pulse shape discrimination performance of stilbene coupled to low-noise silicon photomultipliers
  • Ruch, Marc Lavi; Flaska, Marek; Pozzi, Sara A.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 793 https://doi.org/10.1016/j.nima.2015.04.053
journal September 2015
A study of light collection efficiency in scintillation detectors journal December 1984
Geant4—a simulation toolkit
  • Agostinelli, S.; Allison, J.; Amako, K.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 506, Issue 3 https://doi.org/10.1016/S0168-9002(03)01368-8
journal July 2003
Improvement of the Light Collection in Scintillation Detectors journal January 1979
Digital pulse processing for NaI(Tl) detectors
  • Di Fulvio, A.; Shin, T. H.; Hamel, M. C.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 806 https://doi.org/10.1016/j.nima.2015.09.080
journal January 2016
Resolution and Cathode Uniformity in Scintillation Counters journal January 1967
Measurement of Compton edge position in low-Z scintillators journal March 2010
Contribution of photocathode nonuniformity to energy resolution in NaI(Tl) scintillation detectors
  • Fraser-Mitchell, J.; Wright, A. G.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 288, Issue 2-3 https://doi.org/10.1016/0168-9002(90)90134-R
journal March 1990
Dual-particle imaging system based on simultaneous detection of photon and neutron collision events
  • Poitrasson-Rivière, Alexis; Hamel, Michael C.; Polack, J. Kyle
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 760 https://doi.org/10.1016/j.nima.2014.05.056
journal October 2014
Characterization and calibration of large-volume PVT detectors by Backscatter Gating
  • Pausch, Guntram; Roemer, Katja; Herbach, Claus-Michael
  • 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (2011 NSS/MIC), 2011 IEEE Nuclear Science Symposium Conference Record https://doi.org/10.1109/NSSMIC.2011.6154452
conference October 2011

Similar Records

Effects of Packaging SrI2(Eu) Scintillator Crystals
Journal Article · Sat Jan 01 00:00:00 EST 2011 · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1798640

Time Resolution Measurements of EJ-232Q With Single- and Dual-Sided Readouts
Journal Article · Mon Jul 20 00:00:00 EDT 2020 · IEEE Transactions on Nuclear Science · OSTI ID:1798640

Position Sensitive Photomultiplier Tubes for scintillation imaging
Conference · Sat Feb 01 00:00:00 EST 1986 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:1798640