WHOLE-BODY COUNTER FOR INTERNAL CONTAMINATION CONTROL
Journal Article
·
· Nukleonika (Poland)
OSTI ID:4079485
A facility for body radioactivity measurements was built and put in service in the Institute of Nuclear Research. Two scintillation counters NaI(Tl) working in a coupled detection system were used as detectors. A calibration in vivo for isotope K/sup 42/ was performed. A calibration with phantom was made for Au/sup 198/, Mn/sup 56/, K/sup 42/, and Na/sup 24/. By comparison of the energy spectra obtained for patient and for phantom it was established that phantom gives a good approximation of absorption and scattering of the radiation in the body. The dependence of the sensitivity on radiation energy was established. The K/sup 40/ and Cs/sup 137/ contents in the human body were determined. (auth)
- Research Organization:
- Inst. of Nuclear Research, Polish Academy of Sciences, Warsaw
- NSA Number:
- NSA-18-014260
- OSTI ID:
- 4079485
- Journal Information:
- Nukleonika (Poland), Journal Name: Nukleonika (Poland) Vol. Vol: 8; ISSN NUKLA
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
Semi-Portable Whole-Body Counter for Cesium 137 and Other Gamma-Emitting Isotopes
MEASUREMENTS WITH A WHOLE BODY COUNTER