CERENKOV SPECTROMETER FOR MEASURING GAMMA-RAY ENERGIES
Journal Article
·
· Nucl. Instr. Methods
The calibration experiment is described of a full absorption Cerenkov counter made of lead glass for energy measurements of gamma quanta and electrons with energies of several Bev. The counter has been tested by monoenergetic electrons within the energy range between 0.5 and 3.0 Bev. The Cerenkov counter pulse amplitude versus electron energy follows a linear law within the indicated energy range. The energy resolution of the spectrometer, determined as the ratio of pulse amplitude distribution curve half-width to pulse amplitude at the distribution maximum, for 1.0 Bev electrons has been measured to be 29%. (auth)
- Research Organization:
- Joint Inst. for Nuclear Research, Dubna, USSR
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-18-018212
- OSTI ID:
- 4034379
- Journal Information:
- Nucl. Instr. Methods, Journal Name: Nucl. Instr. Methods Vol. Vol: 25
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
CHERENKOV SPECTROMETER FOR MEASURING GAMMA ENERGY
Total Absorption Lead Glass Cerenkov Counter
A Ĩerenkov counter as high energy photon spectrometer
Journal Article
·
Thu Oct 31 23:00:00 EST 1963
· Pribory i Tekhn. Eksperim.
·
OSTI ID:4066403
Total Absorption Lead Glass Cerenkov Counter
Journal Article
·
Sat Dec 31 23:00:00 EST 1960
· Review of Scientific Instruments
·
OSTI ID:4837401
A Ĩerenkov counter as high energy photon spectrometer
Journal Article
·
Sun Sep 01 00:00:00 EDT 1957
· Nuovo Cimento
·
OSTI ID:4353485