Performance characteristics of an S-600 portable atomic absorption spectrophotometer
Journal Article
·
· Industrial Laboratory
OSTI ID:258500
- Elektron Co., Sumy (Russian Federation)
Performance characteristics of an S-600 portable atomic absorption spectrophotometer are discussed. The optimum analysis conditions, characteristic mass, and detection limit for determining Be, Cd, Co, Cr, Cu, Fe, Mn, Ni, Pb, and Zn in solutions and in powders are specified. Direct analysis of solid-state samples (standard soil samples) is described. The relative error of measurement by the calibration graph method lies, with few exceptions, within 7-30%, and by the addition method, within 4-20%. The time needed for a single element determination is 10-20 min.
- OSTI ID:
- 258500
- Journal Information:
- Industrial Laboratory, Journal Name: Industrial Laboratory Journal Issue: 6 Vol. 61; ISSN 0019-8447; ISSN INDLAP
- Country of Publication:
- United States
- Language:
- English
Similar Records
Determination of traces of lead and cadmium in nonferrous metals and their alloys by atomic absorption spectrometry using a graphite atomizer
Trace elements in coal and coke ash by atomic absorption
AN ATOMIC ABSORPTION SPECTROPHOTOMETER
Journal Article
·
Thu May 01 00:00:00 EDT 1986
· Ind. Lab. (Engl. Transl.); (United States)
·
OSTI ID:5338306
Trace elements in coal and coke ash by atomic absorption
Journal Article
·
Sat Dec 31 23:00:00 EST 1977
· Annu. Book ASTM Stand.; (United States)
·
OSTI ID:5651225
AN ATOMIC ABSORPTION SPECTROPHOTOMETER
Journal Article
·
Sun Sep 01 00:00:00 EDT 1963
· Applied Optics (U.S.)
·
OSTI ID:4055690