Effect of Temperature on Thrombogenicity Testing of Biomaterials in an In Vitro Dynamic Flow Loop System

Journal Article · · ASAIO Journal

To develop and standardize a reliablein vitrodynamic thrombogenicity test protocol, the key test parameters that could impact thrombus formation need to be investigated and understood. In this study, we evaluated the effect of temperature on the thrombogenic responses (thrombus surface coverage, thrombus weight, and platelet count reduction) of various materials using anin vitroblood flow loop test system. Whole blood from live sheep and cow donors was used to assess four materials with varying thrombogenic potentials: negative-control polytetrafluoroethylene (PTFE), positive-control latex, silicone, and high-density polyethylene (HDPE). Blood, heparinized to a donor-specific concentration, was recirculated through a polyvinyl chloride tubing loop containing the test material at room temperature (22–24°C) for 1 hour, or at 37°C for 1 or 2 hours. The flow loop system could effectively differentiate a thrombogenic material (latex) from the other materials for both test temperatures and blood species (p< 0.05). However, compared with 37°C, testing at room temperature appeared to have slightly better sensitivity in differentiating silicone (intermediate thrombogenic potential) from the relatively thromboresistant materials (PTFE and HDPE,p< 0.05). These data suggest that testing at room temperature may be a viable option for dynamic thrombogenicity assessment of biomaterials and medical devices.

Research Organization:
Oak Ridge Institute for Science and Education (ORISE), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
SC0014664
OSTI ID:
2425650
Journal Information:
ASAIO Journal, Journal Name: ASAIO Journal Journal Issue: 6 Vol. 69; ISSN 1058-2916
Publisher:
American Society for Artificial Internal Organs
Country of Publication:
United States
Language:
English

References (32)

Thrombogenicity Testing of Medical Devices in a Minimally Heparinized Ovine Blood Loop journal May 2017
Preclinical Evaluation of the Thrombogenicity and Endothelialization of Bare Metal and Surface-Coated Neurovascular Stents journal September 2014
An In Vitro Blood Flow Loop System for Evaluating the Thrombogenicity of Medical Devices and Biomaterials journal February 2020
Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes journal November 2004
A Systematic Evaluation of the Effect of Temperature on Coagulation Enzyme Activity and Platelet Function journal June 2004
Effect of Shear Stress on Platelet Adhesion to Expanded Polytetrafluoroethylene, a Silicone Sheet, and an Endothelial Cell Monolayer journal November 2000
Comparison of catheter thrombogenicity in a modified chandler loop model using goat blood journal September 2018
Room Temperature Activates Human Blood Platelets journal April 2001
Hemocompatibility testing according to ISO 10993-4: Discrimination between pyrogen- and device-induced hemostatic activation journal September 2014
Effect of temperature on platelet adherence journal January 2016
Neurovascular Rescue journal June 2005
Effects of Low Temperature on Shear-Induced Platelet Aggregation and Activation journal August 2004
Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model journal September 2018
Bleeding and Thrombosis in Patients With Continuous-Flow Ventricular Assist Devices journal June 2012
Guide wire thrombus formation during trans-femoral TAVI journal September 2014
Surface Modification of Biomaterials: A Quest for Blood Compatibility journal January 2012
Characterization of the competing role of surface-contact and shear stress on platelet activation in the setting of blood contacting devices journal April 2021
Adhesive Properties of Platelets from Different Animal Species journal February 2003
Biomaterial testing for covered stent membranes: Evaluating thrombosis and restenosis potential journal September 2011
The Effect of Temperature on Blood Coagulation time journal December 1916
Shear Stress‐Induced Total Blood Trauma in Multiple Species journal July 2017
An in vitro model for preclinical testing of thrombogenicity of resorbable metallic stents journal October 2014
Frequency and predictors of thrombus inside the guiding catheter during interventional procedures: an optical coherence tomography study journal October 2014
Hemocompatibility evaluation of different silver nanoparticle concentrations employing a modified Chandler-loop in vitro assay on human blood journal July 2013
A novel approach to extraction of a large thrombus on the intraventricular guide‐wire during transcatheter aortic valve replacement journal September 2015
In Vitro Thrombogenicity Testing of Artificial Organs journal September 1998
Nonthrombogenic Approaches to Cardiovascular Bioengineering journal August 2011
In vitro hemocompatibility studies of drug-loaded poly-(l-lactic acid) fibers journal December 2003
The optimal incubation time for in vitro hemocompatibility testing: Assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions
  • Blok, Sjoerd Leendert Johannes; van Oeveren, Willem; Engels, Gerwin Erik
  • Journal of Biomedical Materials Research Part B: Applied Biomaterials, Vol. 107, Issue 7 https://doi.org/10.1002/jbm.b.34326
journal January 2019
Comparison of Modified Chandler, Roller Pump, and Ball Valve Circulation Models forIn VitroTesting in High Blood Flow Conditions: Application in Thrombogenicity Testing of Different Materials for Vascular Applications journal January 2012
Thrombogenicity testing of central venous cathetersin vitro journal January 1999
Sheep, pig, and human platelet-material interactions with model cardiovascular biomaterials journal June 1999

Similar Records

Comparison of animal and human blood for in vitro dynamic thrombogenicity testing of biomaterials
Journal Article · Sat Jul 30 00:00:00 EDT 2022 · Artificial Organs · OSTI ID:2425618

In Vitro Thrombogenicity Testing of Biomaterials in a Dynamic Flow Loop: Effects of Length and Quantity of Test Samples
Journal Article · Tue Jul 25 00:00:00 EDT 2023 · Journal of Medical Devices · OSTI ID:2425710

Reduced thrombogenic characteristics of expanded polytetrafluoroethylene and polyurethane arterial grafts after heparin bonding
Journal Article · Sat Dec 31 23:00:00 EST 1983 · Surgery; (United States) · OSTI ID:6221561