Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Comparison of animal and human blood for in vitro dynamic thrombogenicity testing of biomaterials

Journal Article · · Artificial Organs
DOI:https://doi.org/10.1111/aor.14366· OSTI ID:2425618
Abstract Background

To determine suitable alternatives to human blood for in vitro dynamic thrombogenicity testing of biomaterials, four different animal blood sources (ovine, bovine, and porcine blood from live donors, and abattoir porcine blood) were compared to fresh human blood.

Methods

To account for blood coagulability differences between individual donors and species, each blood pool was heparinized to a donor‐specific concentration immediately before testing in a dynamic flow loop system. The target heparin level was established using a static thrombosis pre‐test. For dynamic testing, whole blood was recirculated at room temperature for 1 h at 200 ml/min through a flow loop containing a single test material. Four materials with varying thrombotic potentials were investigated: latex (positive control), polytetrafluoroethylene (PTFE) (negative control), silicone (intermediate thrombotic potential), and high‐density polyethylene (HDPE) (historically thromboresistant). Thrombus weight and surface area coverage on the test materials were quantified, along with platelet count reduction in the blood.

Results

While donor‐specific heparin levels varied substantially from 0.6 U/ml to 7.0 U/ml among the different blood sources, each source was able to differentiate between the thrombogenic latex and the thromboresistant PTFE and HDPE materials (p < 0.05). However, only donor ovine and bovine blood were sensitive enough to differentiate an increased response for the intermediate thrombotic silicone material compared to PTFE and HDPE.

Conclusions

These results demonstrated that multiple animal blood sources (particularly donor ovine and bovine blood) may be suitable alternatives to fresh human blood for dynamic thrombogenicity testing when appropriate control materials and donor‐specific anticoagulation levels are used.

Research Organization:
Oak Ridge Institute for Science and Education (ORISE), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
SC0014664
OSTI ID:
2425618
Journal Information:
Artificial Organs, Journal Name: Artificial Organs Journal Issue: 12 Vol. 46; ISSN 0160-564X
Publisher:
Wiley
Country of Publication:
United States
Language:
English

References (31)

Interspecies differences in coagulation profile journal January 2008
Thrombogenicity Testing of Medical Devices in a Minimally Heparinized Ovine Blood Loop journal May 2017
Preclinical Evaluation of the Thrombogenicity and Endothelialization of Bare Metal and Surface-Coated Neurovascular Stents journal September 2014
An In Vitro Blood Flow Loop System for Evaluating the Thrombogenicity of Medical Devices and Biomaterials journal February 2020
Comparison of catheter thrombogenicity in a modified chandler loop model using goat blood journal September 2018
Dacron® vs. PTFE as bypass materials in peripheral vascular surgery – systematic review and meta-analysis journal December 2008
Hemocompatibility testing according to ISO 10993-4: Discrimination between pyrogen- and device-induced hemostatic activation journal September 2014
Comparison of porcine and human coagulation by thrombelastometry journal November 2011
Neurovascular Rescue journal June 2005
Blood Banking and Transfusion Medicine Challenges During the COVID-19 Pandemic journal December 2020
Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model journal September 2018
Bleeding and Thrombosis in Patients With Continuous-Flow Ventricular Assist Devices journal June 2012
Guide wire thrombus formation during trans-femoral TAVI journal September 2014
Adhesive Properties of Platelets from Different Animal Species journal February 2003
Biomaterial testing for covered stent membranes: Evaluating thrombosis and restenosis potential journal September 2011
Shear Stress‐Induced Total Blood Trauma in Multiple Species journal July 2017
An in vitro model for preclinical testing of thrombogenicity of resorbable metallic stents journal October 2014
Haemodilution-induced changes in coagulation and effects of haemostatic components under flow conditions journal December 2013
Species Variability in Platelet Aggregation Response to Different Agonists journal October 2002
On the Use of the Platelet Activity State Assay for the In Vitro Quantification of Platelet Activation in Blood Recirculating Devices for Extracorporeal Circulation journal January 2016
In-vitro study of species-specific coagulation differences in animals and humans using rotational thromboelastometry (ROTEM) journal December 2018
A novel approach to extraction of a large thrombus on the intraventricular guide‐wire during transcatheter aortic valve replacement journal September 2015
In Vitro Thrombogenicity Testing of Artificial Organs journal September 1998
Nonthrombogenic Approaches to Cardiovascular Bioengineering journal August 2011
In Vitro Shear Stress‐Induced Platelet Activation: Sensitivity of Human and Bovine Blood journal June 2013
Preclinical Device Thrombogenicity Assessments: Key Messages From the 2018 FDA, Industry, and Academia Forum journal August 2020
In vitro hemocompatibility studies of drug-loaded poly-(l-lactic acid) fibers journal December 2003
The optimal incubation time for in vitro hemocompatibility testing: Assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions
  • Blok, Sjoerd Leendert Johannes; van Oeveren, Willem; Engels, Gerwin Erik
  • Journal of Biomedical Materials Research Part B: Applied Biomaterials, Vol. 107, Issue 7 https://doi.org/10.1002/jbm.b.34326
journal January 2019
Sheep, pig, and human platelet-material interactions with model cardiovascular biomaterials journal June 1999
Comparing techniques: the use of recalcified plasma in comparison with citrated plasma alone and in combination with thrombin in ultrastructural studies journal November 2011
Platelet Adhesion to Foreign Surfaces Under Controlled Conditions of Whole Blood flow journal January 1977

Similar Records

Effect of Temperature on Thrombogenicity Testing of Biomaterials in an In Vitro Dynamic Flow Loop System
Journal Article · Sat Dec 31 23:00:00 EST 2022 · ASAIO Journal · OSTI ID:2425650

In Vitro Thrombogenicity Testing of Biomaterials in a Dynamic Flow Loop: Effects of Length and Quantity of Test Samples
Journal Article · Tue Jul 25 00:00:00 EDT 2023 · Journal of Medical Devices · OSTI ID:2425710

Reduced thrombogenic characteristics of expanded polytetrafluoroethylene and polyurethane arterial grafts after heparin bonding
Journal Article · Sat Dec 31 23:00:00 EST 1983 · Surgery; (United States) · OSTI ID:6221561

Related Subjects